不同雾化器的功能原理、输送技术和连接方式对气溶胶输送的影响:一项体外研究。

Effect of functional principle, delivery technique, and connection used on aerosol delivery from different nebulizers: An in-vitro study.

机构信息

Clinical Pharmacist at Nasser General Hospital, Beni-Suef, Egypt.

Clinical Pharmacy Department, Faculty of Pharmacy, Beni-Suef University, Beni-Suef, Egypt.

出版信息

Pulm Pharmacol Ther. 2021 Oct;70:102054. doi: 10.1016/j.pupt.2021.102054. Epub 2021 Jul 16.

Abstract

OBJECTIVE

Nebulizers can be divided according to their functional principle into jet, ultrasonic and vibrating mesh nebulizers with intermittent or continuous aerosol delivery and may be used with many different adapters and connections and all can influence their efficiency. The aim of the present work was to evaluate the effect functional principle, delivery technique, and connection used on aerosol delivered from four different nebulizers.

METHODS

Four nebulizers were used in the study; three of them were jet nebulizers (JNs; AeroEclipse, NebuTech, En ful Kit) and one of them was Aerogen Solo vibrating mesh nebulizers (VMN). AeroEclipse and NebuTech are intermittent output nebulizers, while the rest are continuous output nebulizers. Aerogen Solo was used with either a standard T-piece or Aerogen Ultra holding chamber, and En ful Kit was used with a standard T-piece or Circulaire II holding chamber. 2 ml of salbutamol was nebulized to determine the total emitted dose (TED) and aerodynamic droplet characteristics of the emitted aerosol from the 6 different sets (4 nebulizers with T-piece and 2 holding chambers).

RESULTS

The mean ± SD TED from VMN was significantly higher than all the JNs (p < 0.05). Aerogen Ultra with VMN did not show a significant effect on TED compare to T-piece, but it significantly increased (p < 0.05) fine particle dose (FPD; 3091.5 ± 189.4 μg) and fine particle fraction (FPF; 72.7 ± 3.6%). However, the Circulaire II with En ful Kit had significantly higher TED, FPD, and FPF compared to T-piece (p < 0.05). Intermittent JNs (AeroEclipse and NebuTech) had significantly higher TED (p < 0.05) compared to the continuous JNs (En ful Kit) with no significant effect on the other parameters studied. AeroEclipse had the highest MMAD and Aerogen Solo Ultra has the lowest in MMAD.

CONCLUSIONS

The functional principle, delivery technique, and connection used had a significant effect on aerosol delivered from nebulizers. VMNs are significantly better than JNs. Intermittent delivery has significantly better TED than continuous delivery. Holding chamber with both VMNs and JNs improved aerosol delivery compared to standard T-piece.

摘要

目的

根据功能原理,雾化器可分为射流、超声和振动网雾化器,具有间歇或连续气溶胶输送功能,并可与许多不同的适配器和连接装置配合使用,所有这些都可能影响其效率。本研究的目的是评估四种不同雾化器输送的气溶胶的功能原理、输送技术和连接方式的影响。

方法

本研究使用了四种雾化器;其中三种是射流雾化器(JNs;AeroEclipse、NebuTech 和 En ful Kit),一种是 Aerogen Solo 振动网雾化器(VMN)。AeroEclipse 和 NebuTech 是间歇输出雾化器,而其余的是连续输出雾化器。Aerogen Solo 与标准 T 件或 Aerogen Ultra 保持室一起使用,而 En ful Kit 与标准 T 件或 Circulaire II 保持室一起使用。用 2ml 沙丁胺醇雾化,以确定 6 种不同组合(带 T 件的 4 种雾化器和 2 种保持室)的总发射剂量(TED)和发射气溶胶的空气动力学液滴特性。

结果

VMN 的平均±SD TED 明显高于所有 JNs(p<0.05)。与 T 件相比,Aerogen Ultra 与 VMN 联合使用对 TED 没有显著影响,但显著增加(p<0.05)细颗粒剂量(FPD;3091.5±189.4μg)和细颗粒分数(FPF;72.7±3.6%)。然而,与 T 件相比,Circulaire II 与 En ful Kit 联合使用的 TED、FPD 和 FPF 明显更高(p<0.05)。间歇 JNs(AeroEclipse 和 NebuTech)的 TED 明显高于连续 JNs(En ful Kit)(p<0.05),而对其他研究参数无显著影响。AeroEclipse 的 MMAD 最高,Aerogen Solo Ultra 的 MMAD 最低。

结论

功能原理、输送技术和使用的连接方式对雾化器输送的气溶胶有显著影响。VMNs 明显优于 JNs。间歇输送的 TED 明显优于连续输送。VMNs 和 JNs 与标准 T 件相比,使用保持室可改善气溶胶输送。