4 年随访后,骨干骺端锚定压配式干骺端假体下沉:EBRA-FCA 分析。

Subsidence of a metaphyseal-anchored press-fit stem after 4-year follow-up: an EBRA-FCA analysis.

机构信息

Department of Orthopaedics and Traumatology, Medical University of Innsbruck, Anichstrasse 35, 6020, Innsbruck, Austria.

Department of Experimental Orthopaedics, Medical University of Innsbruck, Sonnenburgstr. 16, 6020, Innsbruck, Austria.

出版信息

Arch Orthop Trauma Surg. 2022 Aug;142(8):2075-2082. doi: 10.1007/s00402-021-04068-8. Epub 2021 Jul 21.

Abstract

PURPOSE

Uncemented stem migration analysis by EBRA-FCA (Einzel-Bild-Roentgen Analyse, Femoral Component Analyse) has been seen to be a good predictive indicator for early implant failure. In this study, we investigated the migration behavior of a cementless metaphyseal-anchored press-fit stem after 4-year follow-up.

METHODS

Applying a retrospective study design, we reviewed all consecutive patients who between 2012 and 2017 received a cementless Accolade II press-fit stem at our Department. We reviewed medical histories and performed radiological measurements using EBRA-FCA software. EBRA-FCA measurements and statistical investigations were performed by two independent investigators.

RESULTS

A total of 102 stems in 91 patients (female 60; male 31) fulfilled our inclusion criteria. Mean age at surgery was 66.2 (range 24.3-92.6) years. EBRA migration analysis showed a mean subsidence of 1.4 mm (range 0.0-12.0) at final follow-up. The angle between stem and femur axis was 0.5° (range 0.0°-2.8°) after 48 months. No correlations between gender or Dorr types and subsidence were found (p > 0.05). A body mass index > 30 kg/m showed a significant increase in stem subsidence within the first 6 (p = 0.0258) and 12 months (p = 0.0466) postoperative.

CONCLUSIONS

Migration pattern of the metaphyseal-anchored stem and a low subsidence rate at final follow-up may predict a good long-term clinical result.

TRIAL REGISTRATION

Number: 20181024-1875.

摘要

目的

通过 EBRA-FCA(单幅 X 线分析,股骨部件分析)对非骨水泥柄的迁移分析已被视为早期植入物失败的良好预测指标。在这项研究中,我们研究了一种非骨水泥骨干锚定压配合股骨柄在 4 年随访后的迁移行为。

方法

采用回顾性研究设计,我们回顾了 2012 年至 2017 年期间在我科接受非骨水泥 Accolade II 压配合股骨柄的所有连续患者。我们查阅了病历并使用 EBRA-FCA 软件进行了影像学测量。EBRA-FCA 测量和统计分析由两名独立的研究者进行。

结果

共有 91 例患者(女性 60 例,男性 31 例)的 102 个股骨柄符合我们的纳入标准。手术时的平均年龄为 66.2 岁(范围 24.3-92.6 岁)。EBRA 迁移分析显示最终随访时平均下沉 1.4 毫米(范围 0.0-12.0)。48 个月后,股骨柄与股骨轴之间的角度为 0.5°(范围 0.0°-2.8°)。性别或 Dorr 类型与下沉之间未发现相关性(p>0.05)。体重指数(BMI)>30 kg/m2 在术后 6 个月(p=0.0258)和 12 个月(p=0.0466)时股骨柄下沉明显增加。

结论

骨干锚定股骨柄的迁移模式和最终随访时较低的下沉率可能预示着良好的长期临床结果。

试验注册

注册号:20181024-1875。