Vet Clin Pathol. 2021 Sep;50(3):348-353. doi: 10.1111/vcp.12949. Epub 2021 Jul 20.
Different thromboplastins are available to measure prothrombin time (PT). Stago coagulation analyzers and reagents are currently used in veterinary laboratories and enable PT measurements to explore the coagulation cascade (extrinsic pathway).
The main objective was to compare PT measurements obtained with the STA-NeoPTimal reagent with the commonly used STA-Neoplastine CI Plus reagent. The secondary objective was to compare the PT ratio with the international normalized ratio (INR) calculated from our derived clotting times.
Analytical performance was evaluated with intra-assay and inter-assay precision. Seventy-two individual canine plasma samples were collected. Each sample was tested with both thromboplastins, using an STA Satellite Max analyzer. The PT, PT ratio, and INR values obtained with the two reagents were compared using Passing-Bablok regression for correlations and Bland-Altman plots for method agreements.
The analytical performance of STA-NeoPTimal reagent was acceptable. Compared with the STA-Neoplastine CI Plus reagent, the STA-NeoPTimal reagent showed a positive proportional bias for PT values. Narrow range analyses showed good agreement for normal PT values (less than 9.5 seconds, internal reference cutoff with STA-Neoplastine CI Plus), and clinical concordance was achieved. When PT was prolonged (more than 9.5 seconds), PT increases were more marked with the STA-NeoPTimal reagent. Agreement was good for INR values across the whole range of PT results.
STA-NeoPTimal can be reliably implemented in veterinary laboratories for canine PT measurements, as agreement between the PT results measured with the two reagents was clinically acceptable.
有不同的凝血活酶可供测量凝血酶原时间(PT)。目前,Stago 凝血分析仪和试剂在兽医实验室中得到应用,可用于 PT 测量以探索凝血级联(外源性途径)。
主要目的是比较使用 STA-NeoPTimal 试剂与常用的 STA-Neoplastine CI Plus 试剂获得的 PT 测量值。次要目的是比较 PT 比值与从我们得出的凝血时间计算得出的国际标准化比值(INR)。
用批内和批间精密度评估分析性能。收集了 72 个个体犬血浆样本。使用 STA Satellite Max 分析仪,每个样本均用两种凝血活酶进行检测。使用 Passing-Bablok 回归进行相关性比较和 Bland-Altman 图进行方法一致性比较,比较两种试剂获得的 PT、PT 比值和 INR 值。
STA-NeoPTimal 试剂的分析性能可接受。与 STA-Neoplastine CI Plus 试剂相比,STA-NeoPTimal 试剂的 PT 值呈正比例偏倚。窄范围分析显示,正常 PT 值(小于 9.5 秒,STA-Neoplastine CI Plus 的内部参考截止值)具有良好的一致性,且具有临床一致性。当 PT 延长(大于 9.5 秒)时,STA-NeoPTimal 试剂的 PT 增加更为显著。在整个 PT 结果范围内,INR 值的一致性良好。
STA-NeoPTimal 可在兽医实验室中可靠地用于犬 PT 测量,因为两种试剂测量的 PT 结果之间的一致性具有临床可接受性。