J Cataract Refract Surg. 2021 Aug 1;47(8):1056-1063. doi: 10.1097/j.jcrs.0000000000000583.
To evaluate the comparability and variability of subjective refraction (SR), autorefraction (AR), and wavefront aberrometry (WA) in pseudophakic patients.
Hanusch Hospital, Vienna, Austria.
Prospective study.
Subjective refraction was performed by 2 independent examiners at 2 study visits. Furthermore, 5 measurements of AR and WA were performed in each patient at both visits. Agreement between the 3 refraction methods for spherical equivalent (M) and cylindrical vectors (J0, J45) was analyzed using Friedman multiple comparison and Bland-Altman plots. Predictability of spherical equivalent determined by SR from AR and WA measurements was tested with partial least squares regression and random forest regression.
Ninety eyes of 90 pseudophakic patients were measured. The mean absolute error and arithmetic mean difference of SR measurements of M, J0, and J45 were comparable between both examiners. A small mean difference was found for SR between both visits. Spherical equivalent was more negative when measured by AR [-0.87 diopters (D)] and WA (-0.90 D) compared with SR (-0.60 D), whereas astigmatic vectors agreed well. Good test-retest reliability was found between all 3 refraction methods for M, J0, and J45. Partial least squares regression and random forest regression showed moderate predictive power for M measured by objective refraction and SR.
Reproducibility and reliability of SR measurements in pseudophakic patients showed good agreement. AR and WA measured the spherical equivalent more myopic than SR, whereas astigmatic vectors were comparable between the 3 methods after uneventful cataract surgery.