Veterinary Surgical Center, Vienna, Virginia, USA.
Eastern Pennsylvania Veterinary Medical Center, Allentown, Pennsylvania, USA.
Vet Surg. 2021 Oct;50(7):1389-1397. doi: 10.1111/vsu.13694. Epub 2021 Aug 4.
To evaluate the accuracy of six depth gauges used in three tibial plateau leveling osteotomy (TPLO) plate holes.
Ex vivo experimental study.
Cadaveric canine limbs (n = 10), one 25-mm-thick wood board, and one 33.8-mm-diameter polyvinyl chloride (PVC) pipe.
A TPLO was performed on 10 canine cadaveric pelvic limbs. Three 3.5-mm plate holes were filled with screws. The remaining three plate holes: a compression hole, a combination compression-locking hole, and a stacked combination compression-locking hole were measured by three observers using six commercial depth gauges and using a micrometer as gold standard. The process was repeated on one wood board and one PVC pipe.
Bone measurements collected using two depth gauges with base diameter < 5 mm were smaller than measurements collected using the four depth gauges with base diameter > 5.5 mm (p ranging from < .001 to .038). Mean depth gauge measurements were smaller than micrometer measurements by 2.20 mm for the compression hole, 0.82 mm for the combination hole, and 3.57 mm for the stacked combination hole. Measurement differences among depth gauges were also present for wood board and PVC pipe measurements. Bone measurement variability between depth gauges was less for the combination and compression holes than for the stacked combination hole.
Depth gauges lacked accuracy. Measurements differed among gauges and measurement variability varied based on plate hole geometry.
Depth gauge measurement accuracy varies based on measuring devices and on 3.5-mm plate hole geometry.
评估 6 种胫骨平台水平截骨(TPLO)钢板孔用深度规的准确性。
离体实验研究。
10 只犬尸体四肢,1 块 25mm 厚木板和 1 根 33.8mm 直径聚氯乙烯(PVC)管。
对 10 只犬尸体骨盆肢进行 TPLO 手术。3.5mm 钢板孔填充螺钉。剩余的 3 个钢板孔:压缩孔、组合压缩锁定孔和堆叠组合压缩锁定孔由 3 名观察者使用 6 种商业深度规并用千分尺作为金标准进行测量。该过程在 1 块木板和 1 根 PVC 管上重复进行。
使用基底直径 < 5mm 的 2 种深度规进行骨测量小于使用基底直径 > 5.5mm 的 4 种深度规进行的测量(p 从<.001 到.038)。对于压缩孔,深度规测量值比千分尺测量值小 2.20mm,对于组合孔小 0.82mm,对于堆叠组合孔小 3.57mm。木板和 PVC 管测量也存在深度规之间的测量差异。组合和压缩孔的深度规之间的骨测量变异性小于堆叠组合孔。
深度规缺乏准确性。测量值在不同的量规之间存在差异,且测量变异性基于钢板孔几何形状而变化。
深度规测量精度取决于测量设备和 3.5mm 钢板孔几何形状。