J Shoulder Elbow Surg. 2022 Jan;31(1):201-208. doi: 10.1016/j.jse.2021.06.017. Epub 2021 Aug 2.
Shoulder range-of-motion (ROM) assessment is vital for the follow-up evaluation of operated patients and for the outcome-based research studies. The aim of this study was to investigate the accuracy and reliability of a remote on-screen application (app)-based method of shoulder ROM measurement through a telehealth medium.
A consultant shoulder surgeon, a board-certified orthopedic resident, and a graduate medical doctor served as the examiners. The cohort consisted of 24 healthy volunteers and 16 symptomatic patients with expected shoulder ROM deficits. Shoulder ROM was first examined physically using a goniometer in the clinic and then over Zoom remote conferencing using the protractor extension app of the Chrome browser.
Instrument validity was examined by comparing the goniometer method with the protractor app-based method of the expert shoulder consultant using Bland-Altman analysis. It showed only minor mean differences between the healthy volunteers and the patients in elevation (2.0° and 5.0°, respectively), abduction (2.0° and 3.0°, respectively), external rotation with the elbow adducted (1.9° and 0.2°, respectively), external rotation with the elbow abducted at 90° (0.4° and 4°, respectively), and internal rotation with elbow abducted at 90° (2.3° and 1.2°, respectively), with limits of agreement that were below the well-established minimal clinically important difference values. The intraclass correlation coefficient (ICC) values varied between 0.83 and 0.96 for the volunteers and between 0.90 and 0.98 for the patients, indicating excellent correlation between the 2 methods. The interobserver reliability between 2 examiners for the protractor app-based method as evaluated by ICC scores was excellent; it ranged between 0.86 and 0.98 for the volunteers and between 0.88 and 0.99 for the patients. Comparison of the protractor app-based method with the gold-standard goniometer method for the resident and the graduate doctor showed excellent ICC values.
A protractor app-based method of measuring shoulder ROM over a telehealth medium is accurate and reliable compared with a clinical goniometer method. This validated method can be used during remote telehealth consultation with significant benefits of saving patients travel and time during the COVID-19 (coronavirus disease 2019) pandemic and even later.
肩部活动范围(ROM)评估对于手术患者的随访评估和基于结果的研究至关重要。本研究旨在通过远程医疗手段,研究一种基于远程屏幕应用程序(app)的肩部 ROM 测量方法的准确性和可靠性。
一位顾问肩外科医生、一位认证骨科住院医师和一位研究生医生担任检查者。该队列包括 24 名健康志愿者和 16 名有预期肩部 ROM 缺陷的症状患者。首先在诊所使用量角器进行肩部 ROM 临床检查,然后通过 Zoom 远程会议使用 Chrome 浏览器的量角器扩展应用程序进行检查。
通过比较有经验的肩部顾问使用量角器方法和基于量角器 app 的方法,对仪器的有效性进行了检查,使用 Bland-Altman 分析。它仅显示健康志愿者和患者在抬高(分别为 2.0°和 5.0°)、外展(分别为 2.0°和 3.0°)、肘部内收时外旋(分别为 1.9°和 0.2°)、肘部 90°外展时外旋(分别为 0.4°和 4°)和肘部 90°外展时内旋(分别为 2.3°和 1.2°)方面存在较小的平均差异,一致性界限低于既定的最小临床重要差异值。志愿者的组内相关系数(ICC)值在 0.83 到 0.96 之间,患者的 ICC 值在 0.90 到 0.98 之间,表明两种方法之间具有极好的相关性。通过 ICC 评分评估,两名检查者之间使用量角器 app 进行的观察者间可靠性为优秀;志愿者的 ICC 值范围在 0.86 到 0.98 之间,患者的 ICC 值范围在 0.88 到 0.99 之间。对于住院医师和研究生医生,量角器 app 与金标准量角器方法相比,ICC 值也表现出极好的一致性。
与临床量角器方法相比,一种通过远程医疗手段测量肩部 ROM 的量角器 app 方法准确且可靠。这种经过验证的方法可用于远程远程医疗咨询中,在 COVID-19(冠状病毒病 2019)大流行期间以及以后,可显著为患者节省旅行和时间。