Department of Trauma Surgery, Regensburg University Medical Center, Germany.
Department of Cranio-Maxillofascial Surgery, Regensburg University Medical Center, Germany.
Biomed Res Int. 2021 Jul 26;2021:2321504. doi: 10.1155/2021/2321504. eCollection 2021.
Due to the anatomic structure of the pelvis, free-hand placement of screws in the acetabular fracture management can be difficult. Infra-acetabular screw fixation increases acetabular stability by distal fixation of the cup. Aim of this cadaveric study is to investigate if a plate-referenced drill guide can provide save placement of an infra-acetabular screw over a precontoured suprapectineal quadrilateral buttress plate (SQBP).
We constructed a drill guide for an infra-acetabular screw based on the surface of an anatomically precontoured SQBP. A total of 12 adult cadaveric acetabular specimens were used for drill guide-assisted placement of the infra-acetabular screw. The drill guide contains a radiopaque spiral to allow longitudinal fine adjustment of the SQBP along the pelvic brim to assure correct position of the plate-drill-guide construct in relation to the Koehler's teardrop. After screw placement, we conducted a computed tomography (CT) scan of all specimens to assess the actual position of the screw in relation of the infra-acetabular corridor and the acetabular joint surface.
The position of the screw was within the infra-acetabular corridor in all cases. We did not see any intra-articular or intrapelvic screw penetration. The mean distance of the centerline of the screw to the medial border of the infra-acetabular corridor was 3.35 mm. The secure distance to the virtual surface of the femoral head to was 7.3 mm.
A plate-referenced drill guide can provide safe placement of an infra-acetabular screw for treatment of acetabular fractures. Radiographic fine adjustment is necessary to access the optimal entry point.
由于骨盆的解剖结构,在髋臼骨折处理中徒手放置螺钉可能较为困难。髋臼下螺钉固定通过杯的远端固定增加髋臼稳定性。本尸体研究的目的是研究基于解剖预塑形的四边形前板(SQBP)的钢板参考钻头导向器是否可以安全地放置髋臼下螺钉。
我们根据解剖预塑形的 SQBP 的表面构建了髋臼下螺钉的钻头导向器。总共使用 12 个成人尸体髋臼标本进行辅助钻头导向器放置髋臼下螺钉。钻头导向器包含一个不透射线的螺旋,允许沿骨盆边缘纵向微调 SQBP,以确保钢板-钻头导向器结构相对于 Koehler 泪滴的正确位置。放置螺钉后,我们对所有标本进行计算机断层扫描(CT)扫描,以评估螺钉相对于髋臼下通道和髋臼关节面的实际位置。
在所有情况下,螺钉均位于髋臼下通道内。我们没有看到任何关节内或盆腔内螺钉穿透。螺钉中心线与髋臼下通道内侧边缘的平均距离为 3.35 毫米。与虚拟股骨头表面的安全距离为 7.3 毫米。
钢板参考钻头导向器可安全地放置髋臼下螺钉治疗髋臼骨折。需要进行影像学微调以获得最佳的进入点。