Gaskell S J, Rollins K, Smith R W, Parker C E
Tenovus Institute for Cancer Research, University of Wales College of Medicine, Heath, UK.
Biomed Environ Mass Spectrom. 1987 Dec;14(12):717-22. doi: 10.1002/bms.1200141205.
The determination of serum cortisol by thermospray liquid chromatography/mass spectrometry (LC/MS) has been assessed. The method incorporates stable isotope dilution and immunoadsorption extraction. [M + H]+ ions are monitored during LC/MS. The within-assay reproducibility is satisfactory (coefficient of variation 7% at a concentration of 190 ng ml-1), though inferior to that achieved with the (more lengthy) procedure of gas chromatography/mass spectrometry (GC/MS). Satisfactory agreement between LC/MS (y) and GC/MS (x) data was observed (y = 0.934x + 12.4 ng ml-1; r = 0.968; n = 14). It is concluded that the generation of precise reference data for the assessment of routine cortisol assays is at present better achieved by GC/MS. LC/MS, however, provides satisfactory quantitative data via a more simple and rapid method.