Hiraki Y, Yanagi H, Nagaya I, Shimizu M, Joja I, Aono K
Department of Radiology, Okayama University Medical School, Japan.
Radioisotopes. 1987 Oct;36(10):512-8. doi: 10.3769/radioisotopes.36.10_512.
The authors carried out experimental and clinical studies of left ventricular volume measurement using equilibrium-gated blood pool scan. Using phantoms, we experimentally obtained a regression equation between radioisotope levels and camera counts. From this equation, we deduced a regression equation between camera counts and volumes, which was then used for calculation of left ventricular volumes using count technique. In an experiment using phantom with various volumes, we obtained a favorable result that the average absolute error between the true and measured volumes was as small as 6.8%. In 11 patients, we calculated the left ventricular volume and ejection fraction employing the regression equation and compared the calculated values with the values obtained by cinemascopy. This comparison disclosed an excellent correlation (r = 0.955 for end-diastolic volume, 0.997 for end-systolic volume and 0.853 for ejection fraction). Mean absolute error was 16.2% for end-diastolic volume and 16.0% for end-systolic volume. These results indicate that this technique is clinically applicable.
作者利用平衡门控心血池扫描对左心室容积测量进行了实验和临床研究。通过模型,我们实验性地得出了放射性同位素水平与相机计数之间的回归方程。基于此方程,我们推导了相机计数与容积之间的回归方程,随后该方程被用于采用计数技术计算左心室容积。在使用不同容积模型的实验中,我们获得了良好的结果,即真实容积与测量容积之间的平均绝对误差小至6.8%。在11例患者中,我们采用该回归方程计算了左心室容积和射血分数,并将计算值与电影造影获得的值进行比较。这种比较显示出极佳的相关性(舒张末期容积r = 0.955,收缩末期容积r = 0.997,射血分数r = 0.853)。舒张末期容积的平均绝对误差为16.2%,收缩末期容积的平均绝对误差为16.0%。这些结果表明该技术在临床上是可应用的。