新型系统有效性评估及垂直视频头脉冲试验最佳水平眼位分析。

Evaluation on Effectiveness of a New System as well as Analysis on Optimal Horizontal Eye Position for Vertical Video Head Impulse Test.

机构信息

ENT Institute and Otorhinolaryngology Department of Eye & ENT Hospital, State Key Laboratory of Medical Neurobiology and MOE Frontiers Center for Brain Science, Fudan University, Shanghai, 200031, China.

NHC Key Laboratory of Hearing Medicine (Fudan University), Shanghai, 200031, China.

出版信息

Curr Med Sci. 2021 Aug;41(4):705-711. doi: 10.1007/s11596-021-2415-1. Epub 2021 Aug 17.

Abstract

OBJECTIVE

To compare the performances among three different systems for video head impulse test (vHIT), and to identify an optimal target angle for precisely evaluating the function of vertical semicircular canals in vHIT.

METHODS

A two-center prospective study was done. Participants were sit 1.2 m away from the wall in a noise-proved room that dedicated for vHIT experiments. During the comparison experiments, similar settings were ensured in both hospitals, with the same distance to wall and angle of staring. For each equipment, the procedures followed the developers' recommendations. The same examiner performed the comparison between two systems in one location. For the eye-position projects, targets were placed on the wall sequentially at the pre-marked lines for different angles. For the comparison projects, 9 and 13 participants were recruited, respectively. Any participant with otologic or vestibular disorders was excluded. A total of 26 healthy participants were recruited in the eye-position experiments, 16 of which were further involved in inter-examiner tests.

RESULTS

Our evaluations of three different systems showed that a new vHIT system, VertiGoggles® ZT-VNG-I (VG) performed as good as the long-tested Otometrics® ICS impulse (Oto) and EyeSeeCam® (ESC). During the comparison, we validated 25-degree, instead of right ahead at 0 degree, is a better place to set the targets when torsion was applied at vertical semicircular canal planes.

CONCLUSION

The new VG system is good for clinical practices. Furthermore, we proposed a new protocol to set the targets 25 degrees from right ahead after tilting head 45 degrees to evaluate vertical canals during vHIT.

摘要

目的

比较三种不同的视频头脉冲测试(vHIT)系统的性能,并确定一个最佳的目标角度,以精确评估 vHIT 中垂直半规管的功能。

方法

进行了一项两中心前瞻性研究。参与者坐在距离墙壁 1.2 米的隔音室中进行 vHIT 实验。在比较实验中,两家医院确保了类似的设置,距离墙壁和注视角度相同。对于每种设备,程序都遵循了开发者的建议。同一位检查者在一个地点对两种系统进行了比较。对于眼位项目,目标按照不同角度的预先标记线依次放置在墙上。对于比较项目,分别招募了 9 名和 13 名参与者。任何有耳科或前庭障碍的参与者都被排除在外。共有 26 名健康参与者参与了眼位实验,其中 16 名进一步参与了检查者间测试。

结果

我们对三种不同系统的评估表明,一种新的 vHIT 系统VertiGoggles® ZT-VNG-I(VG)与经过长期测试的 Otometrics® ICS 脉冲(Oto)和 EyeSeeCam®(ESC)表现一样好。在比较过程中,我们验证了在施加垂直半规管平面扭转时,25 度而不是 0 度正前方是设置目标的更好位置。

结论

新的 VG 系统适用于临床实践。此外,我们提出了一种新的方案,即在头倾斜 45 度后将目标设置在正前方 25 度,以在 vHIT 中评估垂直管。