Smith Avery L, Santa Ana Carol A, Fordtran John S, Guileyardo Joseph M
Division of Gastroenterology, Baylor University Medical Center, Dallas, Texas.
Department of Pathology, Baylor University Medical Center, Dallas, Texas.
Proc (Bayl Univ Med Cent). 2021 Jun 16;34(5):560-565. doi: 10.1080/08998280.2021.1930487. eCollection 2021 Sep.
16 Fr Salem Sump™ tubes have special features to facilitate suction drainage of the stomach, including a second lumen for air venting. These tubes are also commonly used to deliver enteral nutrition and medications to intensive care unit (ICU) patients, but we found no previous research to justify this practice. Because of the unused air vent, these tubes have a large external diameter and a small bore infusion channel (no larger than that of a single lumen 12 Fr feeding tube). The causes of 16 Fr Salem Sump tube obstructions in 17 ICU patients included clogged medications (8 cases) and precipitation of feeding formula (7 cases), each of which would be promoted by a narrow bore. Because of multiple drainage holes at their distal end, these tubes cannot be thoroughly cleansed by standard water flushing; moreover, their drainage holes mandate a deeper length of tube insertion beyond the gastroesophageal junction, which increases the likelihood of intestinal or pulmonary perforation. For these reasons, we conclude that 16 Fr Salem Sump tubes are inferior to standard feeding tubes for delivery of enteral nutrition and medications to patients in medical ICUs.
16F Salem Sump™管具有便于胃吸引引流的特殊设计,包括一个用于排气的第二腔。这些管子也常用于向重症监护病房(ICU)患者提供肠内营养和药物,但我们之前未发现有研究能证明这种做法的合理性。由于排气孔未被利用,这些管子外径大而输注通道内径小(不大于单腔12F饲管)。17例ICU患者使用16F Salem Sump管发生堵塞的原因包括药物堵塞(8例)和喂养配方奶沉淀(7例),而狭窄的内径会促使这些情况发生。由于其远端有多个引流孔,这些管子无法通过标准的水洗彻底清洁;此外,其引流孔要求管子插入深度超过胃食管交界处,这增加了肠穿孔或肺穿孔的可能性。基于这些原因,我们得出结论,在向医疗ICU患者提供肠内营养和药物方面,16F Salem Sump管不如标准饲管。