Bera Debabrata, Joshi Radhey Shyam, Majumder Suchit, Mukherjee Sanjeev S
Dept of Cardiology, RTIICS, Kolkata, India.
Dept of Cardiology, RTIICS, Kolkata, India.
Indian Pacing Electrophysiol J. 2022 Jan-Feb;22(1):30-33. doi: 10.1016/j.ipej.2021.09.001. Epub 2021 Sep 4.
A 65-year-old gentleman with dual chamber pacemaker presented with presyncope. The ECG raised concerns of oversensing which was confirmed by magnet response. The device interrogation revealed noise in ventricular channel temporally associated with P wave. The pacing thresholds were normal. Although the ventricular lead impedance was within normal limit the impedance trend suggested 100 Ω decline over last 1 year. RV lead insulation failure was speculated and supported by the bipolar and unipolar intracardiac electrogram. Device was programmed to DOO temporarily. He underwent RV lead replacement uneventfully.
一位植入双腔起搏器的65岁男性患者出现前驱晕厥。心电图引发了对感知过度的担忧,通过磁铁反应得以证实。设备问询显示心室通道存在与P波在时间上相关的噪声。起搏阈值正常。尽管心室导线阻抗在正常范围内,但阻抗趋势表明在过去1年中下降了100Ω。推测右心室导线绝缘故障,并得到双极和单极心内电图的支持。设备被临时程控为DOO模式。他顺利接受了右心室导线更换。