Department of Chemistry, Faculty of Science, King Mongkut's University of Technology Thonburi, Bangkok 10140, Thailand.
Center of Excellence for Innovation in Analytical Science and Technology and Department of Chemistry, Faculty of Sciences, Chiang Mai University, Chiang Mai 50200, Thailand.
Molecules. 2021 Sep 21;26(18):5720. doi: 10.3390/molecules26185720.
A monolithic rod of polyurethane foam-[4-(2-pyridylazo) resorcinol] (PUF-PAR) as a simple chemical sensor for lead assays with smartphone detection and image processing was developed. With readily available simple apparatus such as a plastic cup and a stirrer rod, the monolithic PUF rod was synthesized in a glass tube. The monolithic PUF-PAR rod could be directly loaded by standard/sample solution without sample preparation. A one-shot image in G/B value from a profile plot in ImageJ for a sample with triplicate results via a single standard calibration approach was obtained. A linear single standard calibration was: [G/B value] = -0.038[µg Pb] + 2.827, R = 0.95 for 10-30 µg Pb with a limit of quantitation (LOQ) of 33 µg L. The precision was lower than 15% RSD. The proposed method was tested by an assay for Pb contents in drinking water samples from Bangkok. The results obtained by the proposed method agree with those of ICP-OES and with 100-120% recovery, demonstrating that the method is useful for screening on-site water monitoring.
一种基于聚氨酯泡沫-[4-(2-吡啶偶氮)间苯二酚](PUF-PAR)的整体棒状结构被开发为用于铅分析的简单化学传感器,可与智能手机检测和图像处理配合使用。使用塑料杯和搅拌棒等现成的简单设备,在玻璃管中合成了整体式 PUF 棒。整体式 PUF-PAR 棒可以直接通过标准/样品溶液加载,无需样品制备。通过单标准校准方法,从 ImageJ 中的剖面图获得了带有三重复测试结果的 G/B 值的一次性图像。线性单标准校准为:[G/B 值] = -0.038[µg Pb] + 2.827,R = 0.95,用于 10-30 µg Pb,定量限(LOQ)为 33 µg L。精密度低于 15% RSD。该方法已通过对曼谷市自来水中铅含量的测定进行了测试。所提出的方法的结果与 ICP-OES 的结果一致,回收率为 100-120%,表明该方法可用于现场水质监测的筛选。