Radiat Prot Dosimetry. 2021 Nov 12;196(3-4):248-256. doi: 10.1093/rpd/ncab144.
The study aims to correlate the effective diameter (Deff) and water-equivalent diameter (Dw) parameters with anterior-posterior (AP), lateral (LAT) and AP + LAT dimensions in order to estimate the patient dose in head CT examinations. Seventy-four patient datasets from head CT examinations were retrospectively collected. The patient's sizes were calculated from the middle slice using a software of IndoseCT. Dw and Deff were plotted as functions of AP, LAT and AP + LAT dimensions. The best trendline fit for LAT and AP functions was a second order polynomial, which resulted in R2 of 0.89 for Deff vs LAT, 0.88 for Dw vs LAT, 0.92 for Deff vs AP and 0.91 for Dw vs AP. A linear correlation was found for Deff vs AP + LAT, Dw vs AP + LAT and Dw vs Deff with R2 of 0.97, 0.96 and 0.98, respectively.
本研究旨在将有效直径(Deff)和水等效直径(Dw)参数与前后(AP)、侧向(LAT)和 AP+LAT 尺寸相关联,以便估算头部 CT 检查中的患者剂量。回顾性收集了 74 例头部 CT 检查的患者数据集。使用 IndoseCT 软件从中层计算患者的大小。绘制 Dw 和 Deff 作为 AP、LAT 和 AP+LAT 尺寸的函数。LAT 和 AP 函数的最佳趋势线拟合是二次多项式,Deff 与 LAT 的 R2 为 0.89,Dw 与 LAT 的 R2 为 0.88,Deff 与 AP 的 R2 为 0.92,Dw 与 AP 的 R2 为 0.91。Deff 与 AP+LAT、Dw 与 AP+LAT 和 Dw 与 Deff 之间存在线性相关,R2 分别为 0.97、0.96 和 0.98。