Suppr超能文献

中耳及有源经皮骨传导听力植入装置的患者报告结局

Patient-Reported Outcomes in Middle Ear and Active Transcutaneous Bone Conduction Hearing Implants.

作者信息

E M Jones Stephen, Roplekar-Bance Rujuta, Green Richard, Rae Caroline, Ferguson Aaron, Spielmann Patrick M

机构信息

Department of Otolaryngology, Ninewells Hospital and Medical School, Ninewells Avenue, Dundee, United Kingdom.

Department of Otolaryngology, Addenbrooke's Hospital, Hills Road, Cambridge, United Kingdom.

出版信息

J Int Adv Otol. 2021 Sep;17(5):405-411. doi: 10.5152/iao.2021.21077.

Abstract

OBJECTIVE

This study used questionnaires to examine the patient-reported satisfaction with 2 hearing implant devices to determine the level of overall satisfaction with the devices, which, if any, factors predicted good or poor perceived outcomes, or whether there were any specific aspects of the devices where dissatisfaction was apparent.

METHODS

A post-treatment questionnaire survey of 39 adult patients who had received a Vibrant Soundbridge (VSB) or Bonebridge (BB) hearing implant, with at least 3 months of follow-up, was conducted using the Glasgow Benefit Inventory (GBI) and Hearing Device Satisfaction Scale (HDSS). Satisfaction scores were compared to pre- and post-operative audiologic outcomes. The correlation between GBI and HDSS scores was also examined.

RESULTS

A total of 28 of the 39 patients (72%) responded: 13 with a BB and 15 with a VSB at a mean of 13 months after implantation. The overall mean total GBI score was 30, with no significant differences across the groups. The responders generally reported that they were "satisfied" across most domains of the HDSS. In the study, 25 of the 28 responders were largely satisfied with their devices but 3 respondents were not. Two were known non-users, while one used the device but did not gain the benefit expected. It is instructive to note that all of these dissatisfied recipients were close to the manufacturer recommended limits for implantation of their respective devices at the time of surgery. Certain themes were identified within the patients' responses, indicating common aspects where satisfaction was poorer.

CONCLUSION

This series of 28 implant recipients demonstrates high levels of satisfaction with implantable hearing devices across 2 different validated questionnaires. Implant teams could exercise caution and manage patient expectations if the patients are close to the recommended limits of a particular device.

摘要

目的

本研究通过问卷调查来检验患者报告的对两种听力植入设备的满意度,以确定对这些设备的总体满意度水平,哪些因素(若有)可预测良好或较差的感知结果,以及这些设备是否存在明显令人不满的特定方面。

方法

使用格拉斯哥获益量表(GBI)和听力设备满意度量表(HDSS),对39名接受了活力声桥(VSB)或骨桥(BB)听力植入且至少随访3个月的成年患者进行了治疗后问卷调查。将满意度得分与术前和术后的听力学结果进行比较。还检验了GBI和HDSS得分之间的相关性。

结果

39名患者中有28名(72%)做出了回应:13名使用骨桥,15名使用活力声桥,平均植入后13个月。总体平均GBI总分是30,各组之间无显著差异。回应者普遍报告称,他们在HDSS的大多数领域都“满意”。在该研究中,28名回应者中有25名对其设备基本满意,但有3名回应者不满意。两名是已知的未使用者,而一名使用了该设备但未获得预期益处。值得注意的是,所有这些不满意的接受者在手术时都接近各自设备制造商推荐的植入极限。在患者的回答中确定了某些主题,表明了满意度较低的共同方面。

结论

这组28名植入接受者通过两种不同的有效问卷表明对可植入听力设备的满意度较高。如果患者接近特定设备的推荐极限,植入团队应谨慎行事并管理患者的期望。