Backside polyethylene wear in reverse shoulder arthroplasty.

机构信息

Hospital for Special Surgery, New York, NY, USA.

Hospital for Special Surgery, New York, NY, USA.

出版信息

J Shoulder Elbow Surg. 2022 Mar;31(3):545-552. doi: 10.1016/j.jse.2021.09.002. Epub 2021 Oct 5.

Abstract

BACKGROUND

Aseptic loosening from implant-associated osteolysis in reverse shoulder arthroplasty (RSA) may contribute to premature implant failure. Although articular side polyethylene (PE) damage has been well documented in the literature, no studies to date have investigated backside wear in RSA. The aims of this investigation were to (1) document and compare the damage between the backside and articular surface in explanted RSA components, (2) assess whether certain quadrants have a greater propensity for damage, and (3) report the most common mode(s) of backside PE damage.

METHODS

Twenty-one RSA humeral liners retrieved during revision procedures between 2005 and 2014 were included for analysis. The mean time between implantation and extraction was 16 months (10 days-88 months). Diagnoses at the time of revision included dislocation (10), infection (4), mechanical failure (3), loosening (2), and unknown (2). Liners were examined under light microscopy (×10-30 magnification) and damage on the articular and backside of the liner surface was graded using the modified Hood score. The location and damage modality were compared between the articular side and backside of the implant.

RESULTS

Damage was noted on the articular surfaces of all 21 liners and on the backside surface of 20 liners. The total damage in all the quadrants was higher on the articular surface than on the backside of the component, with a mean difference in total quadrant damage scores of 11.74 ± 3.53 (P < .001). There was no difference in damage among the quadrants on the backside (P = .44) or the articular surface (P = .08). The articular side exhibited greater scratching, abrasion, and surface deformation than the backside (P < .001).

CONCLUSIONS

This short-term retrieval study demonstrated that backside PE damage occurs on the humeral component of RSA implants. There was greater damage to the articular side of the liner but wear to the backside was present in almost all liners. The clinical importance of backside wear in RSA and its overall contribution to PE particulate disease and osteolysis needs further investigation.

摘要

背景

在反式肩关节置换术(RSA)中,与植入物相关的骨溶解引起的无菌性松动可能导致植入物提前失效。尽管关节侧聚乙烯(PE)损伤在文献中已有详细记载,但迄今为止尚无研究调查 RSA 中的背面磨损。本研究的目的是:(1)记录并比较翻修 RSA 组件中关节面和背面的损伤;(2)评估是否某些象限更容易受损;(3)报告背面 PE 损伤的最常见模式。

方法

纳入 2005 年至 2014 年间在翻修手术中取出的 21 个 RSA 肱骨衬垫进行分析。植入和取出之间的平均时间为 16 个月(10 天-88 个月)。翻修时的诊断包括脱位(10 例)、感染(4 例)、机械故障(3 例)、松动(2 例)和原因不明(2 例)。衬垫在光学显微镜下(×10-30 倍放大)进行检查,并使用改良胡德评分对衬垫表面的关节面和背面的损伤进行分级。比较植入物关节面和背面的位置和损伤模式。

结果

21 个衬垫的关节面均有损伤,20 个衬垫的背面表面也有损伤。所有象限的总损伤在关节面都高于组件的背面,关节面和组件背面的总象限损伤评分平均差异为 11.74±3.53(P<.001)。背面(P=.44)或关节面(P=.08)各象限之间的损伤无差异。关节面的划痕、磨损和表面变形比背面更严重(P<.001)。

结论

这项短期取出研究表明,RSA 植入物肱骨组件的背面发生了 PE 损坏。衬垫的关节面损伤更严重,但几乎所有衬垫的背面都有磨损。RSA 背面磨损的临床重要性及其对 PE 颗粒病和骨溶解的总体影响需要进一步研究。