Department of Radiology, Okayama University Medical School, 2-5-1 Shikata-cho, Kita-ku, Okayama, 700-8558, Japan.
Center for Innovative Clinical Medicine, Okayama University Hospital, 2-5-1 Shikata-cho, Kita-ku, Okayama, Japan.
Jpn J Radiol. 2022 Apr;40(4):430-438. doi: 10.1007/s11604-021-01211-8. Epub 2021 Nov 5.
To evaluate the feasibility of percutaneous needle biopsy using a 1.2 Tesla open magnetic resonance imaging (MRI) system, which has the highest field strength among the currently available open MRI systems.
This single-center prospective study included 10 patients. The primary endpoint was the feasibility of biopsy needle insertion into a target lesion under 1.2 Tesla open MRI guidance. The secondary endpoints included adverse events, device failures, and success of tissue specimen acquisition. Biopsy was performed for targets in various organs using an MRI-compatible coaxial needle system consisting of a 16G introducer needle and 18G semi-automatic biopsy needle. A newly developed body coil with a suitable design for intervention was used for intraprocedural imaging.
Biopsy procedures were performed for six musculoskeletal masses, two retroperitoneal masses, one renal mass, and one liver mass. The median diameter of the targets was 4.9 cm (range 2.1-22.8 cm). MRI-guided biopsy needle insertion was feasible in all 10 patients. In total, four grade 1 adverse events (as per Common Terminology Criteria for Adverse Events version 4.0) occurred in three patients. Adequate biopsy specimens for pathological diagnosis were successfully obtained from all 10 patients.
Percutaneous needle biopsy using a 1.2 Tesla open MRI system was feasible for relatively large targets, especially in the musculoskeletal region.
评估在当前可用的开放式磁共振成像(MRI)系统中磁场强度最高的 1.2 特斯拉开放式 MRI 系统进行经皮穿刺活检的可行性。
本单中心前瞻性研究纳入了 10 名患者。主要终点是在 1.2 特斯拉开放式 MRI 引导下将活检针插入目标病变的可行性。次要终点包括不良事件、器械故障和组织标本采集的成功率。使用由 16G 导入针和 18G 半自动活检针组成的兼容 MRI 的同轴针系统对来自不同器官的目标进行活检。使用专为介入设计的新型合适的体线圈进行术中成像。
对 6 个肌肉骨骼肿块、2 个腹膜后肿块、1 个肾肿块和 1 个肝肿块进行了活检程序。目标的中位数直径为 4.9cm(范围 2.1-22.8cm)。10 名患者均可行 MRI 引导下活检针插入。共有 3 名患者发生 4 例 1 级不良事件(根据不良事件通用术语标准 4.0 版)。10 名患者均成功获得了用于病理诊断的足够的活检标本。
1.2 特斯拉开放式 MRI 系统进行的经皮穿刺活检对于较大的目标是可行的,特别是在肌肉骨骼区域。