影响多分支现成腔内移植物治疗胸腹主动脉瘤内脏适用性的因素

Factors Affecting the Visceral Suitability of a Multibranched off-the-Shelf Endograft for the Treatment of Thoraco-Abdominal Aortic Aneurysms.

作者信息

Mazzaccaro Daniela, Avishay Dor, Nano Giovanni

机构信息

Operative Unit of Vascular Surgery, IRCCS Policlinico San Donato, San Donato Milanese, Milan, Italy.

Operative Unit of Vascular Surgery, IRCCS Policlinico San Donato, San Donato Milanese, Milan, Italy.

出版信息

Ann Vasc Surg. 2022 Apr;81:113-120. doi: 10.1016/j.avsg.2021.08.053. Epub 2021 Nov 4.

Abstract

BACKGROUND

To assess the factors affecting visceral suitability of the use of the Zenith T-branch system in a group of patients with thoraco-adbominal aortic aneurysms (TAAAs).

METHODS

Computer tomography angiography (CTA) of patients who presented a TAAA from 01/2015 to 12/2019 were retrospectively examined. Multi-Planar Reconstructions were performed on CTA images to assess the anatomic suitability of the Zenith T-branch in the visceral district. In particular, the branch deviation angle (BDA), and the branch-length were computed for each target vessel.

RESULTS

Fifty-four CTA were examined. In 33.3% of these patients the presence of either a common origin of the superior mesenteric artery and the celiac trunk, or the diameter of 1 or more visceral/renal artery limited the visceral suitability of the device. All patients except 1 (97.9%) fitted the BDA criterion when the graft was placed in a position in which the BDA for the SMA was 5 degrees to the left. The branch-length criteria was met in all patients, except for 1 (97.9%), when the graft was placed in the center of the aorta. The eccentrical placement of the endograft decreased the suitability to 93.7%.

CONCLUSIONS

The Zenith T-branch system can be suitable in the visceral district for about 67% of patients. The target artery diameter was the most limiting criterion. The central location of the graft within the aortic lumen significantly affected the branch-length distance criteria.

摘要

背景

评估影响一组胸腹主动脉瘤(TAAA)患者使用Zenith T型分支系统时内脏适用性的因素。

方法

回顾性检查2015年1月至2019年12月期间出现TAAA的患者的计算机断层血管造影(CTA)。对CTA图像进行多平面重建,以评估Zenith T型分支在内脏区域的解剖适用性。特别是,计算每个目标血管的分支偏转角(BDA)和分支长度。

结果

检查了54例CTA。在这些患者中,33.3%存在肠系膜上动脉和腹腔干共同起源,或1条或多条内脏/肾动脉直径限制了该装置的内脏适用性。当移植物放置在SMA的BDA向左5度的位置时,除1例(97.9%)外的所有患者均符合BDA标准。当移植物放置在主动脉中心时,除1例(97.9%)外的所有患者均符合分支长度标准。腔内移植物的偏心放置使适用性降至93.7%。

结论

Zenith T型分支系统在内脏区域约67%的患者中可能适用。目标动脉直径是最具限制性的标准。移植物在主动脉腔内的中心位置显著影响分支长度距离标准。