八种牙科扫描仪用于牙桥扫描的准确性(精密度和真实性)比较。

comparison of the accuracy (precision and trueness) of eight dental scanners for dental bridge scanning.

作者信息

Vafaee Fariborz, Mohajeri Mahsa, Gholi Mezerji Naser Mohammad, Zadeh Marouf Ebrahim

机构信息

Department of Prosthodontics, Dental Implant Research Center, Dental Faculty, Hamedan University of Medical Sciences, Hamedan, Iran.

Department of Prosthodontics, Dental Research Center, Dentistry Research Institute, School of Dentistry, Tehran University of Medical Sciences, Tehran, Iran.

出版信息

Dent Res J (Isfahan). 2021 Oct 21;18:84. eCollection 2021.

Abstract

BACKGROUND

Dental scanners play a critical role in computer-aided design/computer-aided manufacturing technology. This study aimed to compare the accuracy (precision and trueness) of eight dental scanners for dental bridge scanning.

MATERIALS AND METHODS

In this experimental study, a typodont model with a missing maxillary right first molar was prepared for a 3-unit fixed partial denture. Each scanner (Sirona inEos inLab, Sirona X5, Dentium, Imes icore 350I I3D, Amann Girrbach map 100, 3Shape D100, 3Shape E3) performed seven scans of the typodont, and the data were analyzed using 3D-Tool software. The abutment length, abutment width, arch length, and interdental distance were measured. To assess the accuracy of each scanner, trueness was evaluated by superimposing the scanned data on true values obtained by the 3shape Triosscanner as the reference. Precision was evaluated by superimposing a pair of data sets obtained from the same scanner. Precision and trueness of the scanners were compared using the one-way ANOVA followed by the Tukey's HSD test and one-sample -test (P<0.05 was considerer significant).

RESULTS

The precision of scanners ranged from 14 μm (3Shape Trios) to 45 μm (Imes icore 350i), whereas the trueness ranged from 38 μm (3Shape d700) to 71 μm (Sirona X5).

CONCLUSION

The reported trueness values for 3Shape Trios, Sirona inEos inLab, Sirona x5, Dentium, Imes icore350i, Amann Girrbach, 3Shape d700, and 3Shape e3 were 63, 45, 71, 67, 70, 53, 38, and 42 μm, respectively, whereas the precision values were 14, 29, 44, 34, 45, 44, 30 and 28 μm, respectively.

摘要

背景

牙科扫描仪在计算机辅助设计/计算机辅助制造技术中起着关键作用。本研究旨在比较八款牙科扫描仪在扫描牙桥时的准确性(精度和真值)。

材料与方法

在这项实验研究中,制备了一个上颌右侧第一磨牙缺失的牙模型,用于制作一个三单位固定局部义齿。每台扫描仪(西诺德inEos inLab、西诺德X5、登泰克、艾姆斯iCore 350I I3D、阿曼吉尔巴赫map 100、3Shape D100、3Shape E3)对牙模型进行七次扫描,并使用3D-Tool软件对数据进行分析。测量基牙长度、基牙宽度、牙弓长度和牙间隙距离。为评估每台扫描仪的准确性,通过将扫描数据与以3Shape Trios扫描仪获得的真值进行叠加来评估真值,以3Shape Trios扫描仪作为参考。通过叠加从同一台扫描仪获得的一对数据集来评估精度。使用单因素方差分析,随后进行Tukey's HSD检验和单样本检验来比较扫描仪的精度和真值(P<0.05被认为具有显著性)。

结果

扫描仪的精度范围为14μm(3Shape Trios)至45μm(艾姆斯iCore 350i),而真值范围为38μm(3Shape d700)至71μm(西诺德X5)。

结论

报告的3Shape Trios、西诺德inEos inLab、西诺德x5、登泰克、艾姆斯iCore350i、阿曼吉尔巴赫、3Shape d700和3Shape e3的真值分别为63、45、71、67、70、53、38和42μm,而精度值分别为14、29、44、34、45、44、30和28μm。