使用附加脚环提高六足外固定架参数测量的术后 X 光片质量。

Improving postoperative radiographs for the parameter measurement of hexapod external fixator using an additional foot ring.

机构信息

Department of Trauma and Microreconstructive Surgery, The First Affiliated Hospital of Xinjiang Medical University, Urumqi, Xinjiang, China.

Department of Orthopedics and Trauma, Tianjin Hospital, Tianjin, China.

出版信息

J Orthop Surg Res. 2021 Nov 13;16(1):668. doi: 10.1186/s13018-021-02820-9.

Abstract

BACKGROUND

It is challenging to determine the orthogonality of radiographs in daily clinical practice. The purpose of this study was to show the usefulness of an additional foot ring which might determine the orthogonality of postoperative radiographs for the parameter measurement of hexapod external fixator.

METHODS

We retrospectively analyzed 81 consecutive trauma patients with tibial shaft fractures treated by the hexapod external fixator at our institution from September 2014 to July 2019. Starting in March 2016, the postoperative radiographs for parameter measurement were obtained under the control of an additional foot ring. The final data consisted of 47 patients in traditional radiographs (Group I) and 34 patients under the control of foot ring during the radiographic process (Group II). The demographic data, original postoperative deformities, residual deformities after final correction, number of repeated radiographs after the first postoperative radiographs, time to the satisfactory reduction achieved, and external fixation time in all patients were documented and analyzed. The Johner-Wruhs criteria were used for the final clinical outcomes evaluation at the last clinical visit.

RESULTS

Satisfactory reduction and bone union were achieved in all patients. There were no statistical significances between the two groups in the demographic data, original postoperative deformities, residual deformities after final correction, external fixation time, and the final clinical outcomes (P > 0.05). The mean number of repeated radiographs after the first radiographs (1.4 times) and mean time to the satisfactory reduction achieved (3.3 days) in patients with an additional foot ring used were all less than those without foot ring (2.4 times, 5.3 days) (P < 0.05).

CONCLUSIONS

The additional foot ring is a practical device to ensure the orthogonality of postoperative radiographs for the hexapod external fixator parameter measurement. Radiation exposure, duration of deformity correction, and cost for patients might be reduced due to the less repeated radiographs with the wrong position.

摘要

背景

在日常临床实践中,确定 X 光片的正交性具有挑战性。本研究旨在展示附加足环的实用性,该足环可确定六足外固定器术后 X 光片的正交性,用于参数测量。

方法

我们回顾性分析了 2014 年 9 月至 2019 年 7 月在我院接受六足外固定器治疗的 81 例胫骨骨干骨折连续创伤患者。自 2016 年 3 月起,在附加足环的控制下获取用于参数测量的术后 X 光片。最终数据包括传统 X 光片(I 组)的 47 例患者和 X 光过程中控制足环的 34 例患者(II 组)。记录和分析所有患者的人口统计学数据、原始术后畸形、最终矫正后的残余畸形、第一次术后 X 光片后重复 X 光片的数量、达到满意复位的时间以及外固定时间。在最后一次临床就诊时,使用 Johner-Wruhs 标准对最终临床结果进行评估。

结果

所有患者均获得满意的复位和骨愈合。两组在人口统计学数据、原始术后畸形、最终矫正后残余畸形、外固定时间和最终临床结果方面无统计学差异(P > 0.05)。使用附加足环的患者第一次 X 光片后重复 X 光片的平均次数(1.4 次)和达到满意复位的平均时间(3.3 天)均少于未使用足环的患者(2.4 次,5.3 天)(P < 0.05)。

结论

附加足环是一种实用的设备,可以确保六足外固定器参数测量术后 X 光片的正交性。由于减少了错误位置的重复 X 光片,患者的辐射暴露、畸形矫正持续时间和成本可能会降低。