双层滤过对对比增强能谱乳腺摄影低能图像(LE-CESM)中平均腺体剂量(MGD)和图像质量的影响。

Influence of double layer filter on mean glandular dose (MGD) and image quality in low energy image of contrast enhanced spectral mammography (LE-CESM).

机构信息

Department of Radiological Sciences, Graduate School of Human Health Sciences, Tokyo Metropolitan University, Tokyo, Japan; Department of Radiography & Radiotherapy, Faculty of Allied Health Sciences, General Sir John Kotelawala Defence University, Werahera, Sri Lanka.

Department of Radiological Sciences, Graduate School of Human Health Sciences, Tokyo Metropolitan University, Tokyo, Japan; Department of Radiology, Juntendo University Shizuoka Hospital, Japan.

出版信息

Radiography (Lond). 2022 May;28(2):340-347. doi: 10.1016/j.radi.2021.11.004. Epub 2021 Nov 24.

Abstract

INTRODUCTION

The aim of this study was to evaluate mean glandular dose (MGD) and image quality in low energy imaging from contrast-enhanced spectral mammography (CESM) when using double-layer filtration.

METHODOLOGY

A dedicated phantom was used to quantitatively estimate the MGD and image quality. The target slab of the phantom consisted of three iodine coins having a concentration of 1.0 mgI/cm, 2.0 mgI/cm, 4.0 mgI/cm, a 100% adipose equivalent coin and a 100% glandular equivalent coin. The phantom was exposed using a semiautomated function at 28 k, 30 kV and 32 kV. MGD. Contrast to noise ratio (CNR) and figure of merit (FOM) were estimated for Mo/Rh, Mo/Rh + Cu, Mo/Rh + Al and Mo/Rh + Cd combinations using three breast equivalent compositions.

RESULTS

MGD was reduced up to a maximum of 1.03 mGy from 1.17 mGy for 100% adipose tissue. 1.18 mGy from 1.34 mGy for 50% glandular tissue and 1.39 mGy from 1.72 mGy for the 100% glandular phantom when using double-layer filtration. All of the above-mentioned results were obtained for the 50 mm phantom using 32 kV. CNR and FOM values were not significantly reduced with a double-layer filter when compared to a single-layer filter.

CONCLUSION

The present study concluded that Mo/Rh + Cu is the best combination to reduce the MGD significantly when compared to Mo/Rh + Al or Mo/Rh + Cd. Mo/Rh + Cu also achieved optimal image quality when compared to the Mo/Rh single filter combination.

IMPLICATIONS OF PRACTICE

The use of a double-layer filter in low energy imaging of CESM results in a significant reduction in MGD without degrading the quality of the image.

摘要

简介

本研究旨在评估对比增强能谱乳腺摄影(CESM)中使用双层滤过时低能成像的平均腺体剂量(MGD)和图像质量。

方法

使用专用模体定量估计 MGD 和图像质量。模体的靶板由三个碘币组成,浓度分别为 1.0 mgI/cm、2.0 mgI/cm、4.0 mgI/cm、100%脂肪等效币和 100%腺体等效币。模体在 28 k、30 kV 和 32 kV 下使用半自动功能进行曝光。使用三种乳腺等效组成物,对 Mo/Rh、Mo/Rh+Cu、Mo/Rh+Al 和 Mo/Rh+Cd 组合分别估计 MGD、对比噪声比(CNR)和品质因数(FOM)。

结果

当使用双层滤过时,对于 100%脂肪组织,MGD 从 1.17 mGy 降低到 1.03 mGy,最大降低 1.03 mGy;对于 50%腺体组织,MGD 从 1.34 mGy 降低到 1.18 mGy,最大降低 0.16 mGy;对于 100%腺体模体,MGD 从 1.72 mGy 降低到 1.39 mGy,最大降低 0.33 mGy。所有上述结果均使用 32 kV 和 50mm 模体获得。与单层滤过相比,使用双层滤过时 CNR 和 FOM 值没有明显降低。

结论

本研究得出结论,与 Mo/Rh+Al 或 Mo/Rh+Cd 相比,Mo/Rh+Cu 是降低 MGD 的最佳组合。与 Mo/Rh 单滤过组合相比,Mo/Rh+Cu 还实现了最佳的图像质量。

实践意义

在 CESM 的低能成像中使用双层滤过可显著降低 MGD,而不会降低图像质量。