Dipartimento di Ingegneria, Università Degli Studi di Palermo, Italy.
Istituto di Fisiologia Clinica (IFC)-Consiglio Nazionale Delle Ricerche-Reggio Calabria-Italy, Italy.
Anal Chim Acta. 2022 Jan 15;1190:339215. doi: 10.1016/j.aca.2021.339215. Epub 2021 Oct 27.
In this work electrochemical sensors fabricated from compact disc material (waste or new) are used to quantify chloride ions in different types of samples. All three electrodes, working, counter, and pseudo-reference electrodes, were fabricated from the compact disc and directly used. Different parameters were studied in order to demonstrate the possibility of using this waste material for efficient and low-cost electrochemical sensors. Chloride sensing performance was evaluated using linear scan voltammetry as the detection technique. A sensitivity of 0.174 mA mM cm with a limit of detection of 20 μM and excellent selectivity against many interferents was observed. Selectivity and reproducibility tests were also carried out, showing excellent results. Sensors were also validated with real samples (drinking and sea water, milk, sweat and physiological solutions) with results comparable to conventional techniques. Our results show the applicability and suitability of these low-cost sensors, for detection of those analytes for which, silver, has high sensitivity and selectivity.
在这项工作中,使用由光盘材料(废旧或新的)制成的电化学传感器来定量分析不同类型样品中的氯离子。工作电极、对电极和伪参比电极均由光盘制成,并直接使用。研究了不同的参数,以证明这种废物材料用于高效和低成本电化学传感器的可能性。采用线性扫描伏安法作为检测技术评估了氯离子的传感性能。观察到的灵敏度为 0.174 mA mM cm,检测限为 20 μM,对许多干扰物具有优异的选择性。还进行了选择性和重现性测试,结果非常出色。用实际样品(饮用水和海水、牛奶、汗水和生理溶液)对传感器进行了验证,结果与传统技术相当。我们的结果表明,这些低成本传感器适用于检测那些对银具有高灵敏度和选择性的分析物。