Talanta. 2022 Feb 1;238(Pt 1):122994. doi: 10.1016/j.talanta.2021.122994. Epub 2021 Nov 3.
We developed a small fluorescence microplate reader with an organic photodiode (OPD) array. The OPD array has nine OPDs that have a large light receiving area (9.62 mm per one OPD). Since the OPD array is fabricated on a flat glass plate, it can be placed just below microwells and can detect fluorescence emitted through the entire surface of the microwell bottom. The analytical performance of the developed plate reader was evaluated by measuring an aqueous solution of resorufin. The limit of detection (LOD) for resorufin (0.01-0.05 μM) was lower than that obtained with a plate reader equipped with nine inorganic photodiodes developed in a previous study (0.30 μM) and a commercially available microplate reader (0.16 μM). These results indicate that the large light receiving area improves the detection performance of the system. In addition, the developed reader was successfully used to quantify immunoglobulin A (IgA) in human saliva. The LOD for IgA was estimated to be 1.2 ng/mL, which is low enough to objectively evaluate human stress.
我们开发了一种带有有机光电二极管 (OPD) 阵列的小型荧光微孔板读数仪。该 OPD 阵列有九个 OPD,具有较大的受光面积(每个 OPD 为 9.62 毫米)。由于 OPD 阵列是在平板玻璃上制造的,因此它可以放置在微孔下方,并且可以检测通过微孔底部整个表面发射的荧光。通过测量溶于水中的 Resorufin 来评估所开发的板读数仪的分析性能。Resorufin 的检测限(LOD)(0.01-0.05 μM)低于之前研究中开发的带有九个无机光电二极管的板读数仪(0.30 μM)和市售微孔板读数仪(0.16 μM)的检测限。这些结果表明,较大的受光面积提高了系统的检测性能。此外,该开发的读取器成功用于定量人唾液中的免疫球蛋白 A(IgA)。IgA 的 LOD 估计为 1.2ng/mL,足以客观评估人的压力。