Department of Orthopaedic Surgery, Mie University, Graduate School of Medicine, 2-174 Edobashi, Tsu, Mie, 514-8507, Japan.
BMC Musculoskelet Disord. 2021 Dec 6;22(1):1021. doi: 10.1186/s12891-021-04906-1.
Leg length discrepancy is one of the most common problems after total hip arthroplasty (THA). The aim of this study was to investigate the accuracy of image-free navigation in intraoperative leg length change (LLC) using evaluations from anteroposterior radiographs (2D measurement) and 3D bone models using CT data (3D measurement).
One hundred THAs with cementless cups and stems were performed using an image-free navigation system in our hospital. We evaluated the accuracy of image-free navigation based on LLC from 2D and 3D measurements. Furthermore, we also investigated error in absolute value and correlations between 2D and 3D measurements in LLC.
The accuracy of image-free navigation based on 2D measurement was 94% within 5 mm and 76% within 3 mm. The accuracy of image-free navigation based on 3D measurement was 92% within 5 mm and 81% within 3 mm. The error in absolute value in LLC between 2D and 3D measurements was 1.7 ± 1.4 mm (range, 0 to 6 mm). A strong correlation was observed between 2D and 3D measurements in the LLC.
In the present study, good accuracy of image-free navigation in intraoperative LLC was confirmed for both evaluation methods from 2D and 3D measurements. In addition, the error in absolute value in the LLC between 2D and 3D measurements was very small, and we observed a strong correlation between 2D and 3D measurements. Based on these results, evaluation of LLC from radiographs was considered sufficient if radiographs can be taken accurately.
肢体长度差异是全髋关节置换术后(THA)最常见的问题之一。本研究旨在通过对前后位 X 线片(2D 测量)和 CT 数据的 3D 骨模型进行评估,研究术中肢体长度变化(LLC)的无图像导航的准确性。
我院对 100 例采用无图像导航系统的非骨水泥髋臼杯和柄进行 THA。我们根据 LLC 的 2D 和 3D 测量结果评估无图像导航的准确性。此外,我们还研究了 LLC 中 2D 和 3D 测量之间的绝对值误差和相关性。
基于 2D 测量的无图像导航的准确性为 94%,在 5mm 以内,76%在 3mm 以内。基于 3D 测量的无图像导航的准确性为 92%,在 5mm 以内,81%在 3mm 以内。2D 和 3D 测量 LLC 之间的绝对值误差为 1.7±1.4mm(范围为 0 至 6mm)。2D 和 3D 测量 LLC 之间存在很强的相关性。
在本研究中,通过 2D 和 3D 测量对术中 LLC 的无图像导航的准确性进行了良好的确认。此外,2D 和 3D 测量 LLC 之间的绝对值误差非常小,并且观察到 2D 和 3D 测量之间存在很强的相关性。基于这些结果,如果可以准确拍摄 X 线片,则认为 X 线片对 LLC 的评估是足够的。