Lommen Julian, Schorn Lara, Nitschke Julia, Sproll Christoph, Zeller Uwe, Kübler Norbert R, Handschel Jörg, Holtmann Henrik
Department of Oral-, Maxillo- and Plastic Facial Surgery, Heinrich-Heine-University, Moorenstraße 5, 40225 Duesseldorf, Germany.
Department of Otolaryngology, Kliniken Maria Hilf, Viersener Straße 450, 41063 Mönchengladbach, Germany.
Saudi Dent J. 2021 Dec;33(8):884-889. doi: 10.1016/j.sdentj.2021.09.018. Epub 2021 Sep 20.
The benefits of portable dental X-ray devices remain controversially debated. This study aimed to compare the accuracy in positioning dental X-ray images using handheld (Nomad Pro 2) and wall-mounted (Heliodent Plus) X-ray devices.
Radiographical imaging was exercised on a maxillofacial phantom using the handheld dental X-ray device Nomad Pro 2 (Kavo Kerr, Biberach, Germany) and the wall-mounted dental X-ray device Heliodent Plus (Sirona Dental Systems, Bensheim, Germany). Accuracy of device positioning (i.e., centeredness and perpendicularity) was measured as horizontal and vertical deviation (pixels and millimeters) from a centrally positioned crosshair. The reproducibility of the results was tested for 80 images per device and operator. IBM SPSS (SPSS Inc., Chicago, IL) was used for statistical analyses.
Centeredness of produced dental X-rays was significantly better for the wall-mounted device than for the handheld device in both deviation from the x-axis (p = 0.042) and y-axis (p = 0.020). The perpendicularity of the produced dental X-rays was significantly better for the handheld device than for the wall-mounted device for both horizontal (p < 0.001) and vertical (p < 0.001) plains.
Handheld dental X-ray devices appear to provide a high degree of accuracy in image positioning, especially in regard to proper perpendicular image angulation.
便携式牙科X射线设备的益处仍存在争议。本研究旨在比较使用手持式(Nomad Pro 2)和壁挂式(Heliodent Plus)X射线设备定位牙科X射线图像的准确性。
使用手持式牙科X射线设备Nomad Pro 2(德国比伯拉赫卡瓦科尔公司)和壁挂式牙科X射线设备Heliodent Plus(德国本斯海姆西诺德牙科系统公司)对颌面模型进行放射成像。设备定位的准确性(即中心性和垂直度)通过与位于中心的十字准线的水平和垂直偏差(像素和毫米)来测量。对每个设备和操作员的80张图像进行结果再现性测试。使用IBM SPSS(美国伊利诺伊州芝加哥SPSS公司)进行统计分析。
在X轴偏差(p = 0.042)和Y轴偏差(p = 0.020)方面,壁挂式设备产生的牙科X射线的中心性明显优于手持式设备。在水平(p < 0.001)和垂直(p < 0.001)平面上,手持式设备产生的牙科X射线的垂直度明显优于壁挂式设备。
手持式牙科X射线设备在图像定位方面似乎具有高度准确性,尤其是在正确的垂直图像角度方面。