A computer based simulation method was developed to assess the relative effectiveness and availability of various attenuation compensation algorithms in single photon emission computed tomography (SPECT). The effect of the nonuniformity of attenuation coefficient distribution in the body, the errors in determining a body contour and the statistical noise on reconstruction accuracy and the computation time in using the algorithms were studied. The algorithms were classified into three groups: precorrection, post correction and iterative correction methods. Furthermore, a hybrid method was devised by combining several methods. This study will be useful for understanding the characteristics, limitations and strengths of the algorithms and searching for a practical correction method for photon attenuation in SPECT.