计算机断层扫描噪声和组织异质性对氙增强计算机断层扫描测定脑血流量的影响。

The effect of computed tomography noise and tissue heterogeneity on cerebral blood flow determination by xenon-enhanced computed tomography.

作者信息

Good W F, Gur D

出版信息

Med Phys. 1987 Jul-Aug;14(4):557-61. doi: 10.1118/1.596067.

Abstract

The errors associated with derivation of cerebral blood flow values by the xenon-enhanced computed tomography (CT) method have been evaluated as a function of tissue heterogeneity and CT noise. The results of this study indicate that CT noise introduces large errors in the derived flow value when data for a single, unprocessed voxel are used for this purpose. CT noise increases the derived flow values in a systematic way. Tissue heterogeneity results in a systematic error which lowers the derived flow values. Errors due to both parameters are computed for typical and extreme conditions.

摘要

通过氙增强计算机断层扫描(CT)方法推导脑血流量值时所涉及的误差,已作为组织异质性和CT噪声的函数进行了评估。本研究结果表明,当为此目的使用单个未处理体素的数据时,CT噪声会在推导的流量值中引入较大误差。CT噪声以一种系统的方式增加推导的流量值。组织异质性会导致系统误差,从而降低推导的流量值。针对典型和极端情况计算了由这两个参数引起的误差。