Comparison of the effects of varying tube voltage and iodinated concentration on increasing the iodinated radiation dose in computed tomography.

机构信息

Graduate School of Health Sciences, Kumamoto University, Kumamoto, Japan; Division of Radiology, Okayama University Hospital, Okayama, Japan.

Department of Diagnostic Radiology, Faculty of Life Sciences, Kumamoto University, Kumamoto, Japan.

出版信息

Phys Med. 2022 Mar;95:57-63. doi: 10.1016/j.ejmp.2022.01.007. Epub 2022 Jan 29.

Abstract

PURPOSE

To compare the effects of tube voltage and iodinated concentration on increasing the iodinated radiation dose in computed tomography (CT).

MATERIAL AND METHODS

Water and iodinated materials were inserted in an anthropomorphic thorax phantom. Helical CT scans were performed with tube voltages of 80 and 120 kV. Monte Carlo simulation of radiation doses in CT images was used to study the radiation dose profiles. The radiation doses at different iodine to water ratios in images were calculated from cumulative radiation doses for iodine and water at each iodine concentration, tube voltage, and peripheral/central location. A radiation dose ratio from 80 to 120 kV was calculated from cumulative radiation doses at the same iodine concentration.

RESULTS

The iodinated radiation doses with small and large phantoms were 1.56-2.04 and 1.61-1.82 times higher at 80 kV and 1.55-2.23 and 1.22-1.79 times higher at 120 kV than those in water. In the central portion, the iodinated radiation dose ratio decreased by 1.14-0.75 and 1.20-0.93 times at 80 kV and 1.29-1.23 and 1.17-1.23 times at 120 kV with increasing iodinated concentrations. In the peripheral portion, the iodinated radiation dose was slightly higher for 80 kV than for 120 kV. In the central portion, the ratio decreased with increasing iodinated concentration.

CONCLUSION

The increase in iodinated radiation dose caused by photoelectric absorption was greater with increased iodine load than with lower tube voltage.

摘要

目的

比较管电压和碘浓度对 CT 中碘辐射剂量增加的影响。

材料与方法

将水和含碘材料插入人体胸腔模型。使用 80kV 和 120kV 管电压进行螺旋 CT 扫描。采用蒙特卡罗模拟方法对 CT 图像中的辐射剂量分布进行研究。根据每个碘浓度、管电压和外周/中心位置处的碘和水的累积辐射剂量,计算出图像中不同碘/水比例的辐射剂量。从相同碘浓度的累积辐射剂量计算出 80-120kV 的辐射剂量比。

结果

小模型和大模型的含碘辐射剂量分别在 80kV 时比水高 1.56-2.04 倍和 1.61-1.82 倍,在 120kV 时比水高 1.55-2.23 倍和 1.22-1.79 倍。在中心区域,80kV 时碘浓度增加导致碘辐射剂量比降低 1.14-0.75 倍和 1.20-0.93 倍,120kV 时碘浓度增加导致碘辐射剂量比降低 1.29-1.23 倍和 1.17-1.23 倍。在外周区域,80kV 时的碘辐射剂量略高于 120kV。在中心区域,碘辐射剂量随碘浓度的增加而降低。

结论

光电吸收引起的碘辐射剂量增加随碘负荷的增加而大于管电压的降低。