The Heart Institute, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA.
Department of Pediatrics, University of Cincinnati College of Medicine, Cincinnati, OH, USA.
Pediatr Cardiol. 2022 Aug;43(6):1205-1213. doi: 10.1007/s00246-022-02840-0. Epub 2022 Feb 6.
We evaluate the validity of cardiac index (CI) measurements utilizing the Ultrasonic Cardiac Output Monitor (USCOM), a non-invasive Doppler ultrasound device, by comparing measurements to cardiac catheterization-derived CI measurements in patients with single-ventricle physiology. USCOM measurements were repeated three times for each patient at the beginning of a cardiac catheterization procedure for twenty-six patients undergoing elective pre-Glenn or pre-Fontan catheterization. CI was measured by USCOM and was calculated from cardiac catheterization data using Fick's method. Bland-Altman analysis for CI showed bias of 0.95 L/min/m with the 95% limits of agreement of - 1.85 and 3.75. Pearson's correlation coefficient was 0.89 (p < 0.001) indicating a strong positive relationship between USCOM and cardiac catheterization CI measurements. When excluding two patients with significant dilation of the neo-aortic valve (z-score > + 5), the bias improved to 0.66 L/min/m with the 95% limits of agreement of - 1.38 and 2.70. Percent error of limits of agreement was 34%. There was excellent intra-operator reproducibility of USCOM CI measurements with an intra-class coefficient of 0.96. We demonstrate the use of USCOM to measure CI in patients with single-ventricle physiology for the first time, showing acceptable agreement of the CI measurements between USCOM and cardiac catheterization with a high intra-operator reproducibility.
我们通过比较超声心动图(USCOM)和心导管测量的心输出量指数(CI),评估了一种非侵入性多普勒超声设备测量心脏指数的有效性,该研究对象为单心室生理患者。26 例行择期 Glenn 或 Fontan 心导管检查的患者在心导管检查前进行了三次 USCOM 测量,每次测量均重复三次。USCOM 通过 Fick 法从心导管数据中计算 CI。CI 的 Bland-Altman 分析显示,USCOM 的偏差为 0.95 L/min/m,95%一致性区间为-1.85 和 3.75。Pearson 相关系数为 0.89(p<0.001),表明 USCOM 和心导管 CI 测量之间存在强正相关。当排除两名升主动脉瓣显著扩张的患者(z 分数> + 5)后,偏差改善至 0.66 L/min/m,95%一致性区间为-1.38 和 2.70。一致性界限的百分比误差为 34%。USCOM CI 测量的操作者内可重复性极好,组内相关系数为 0.96。我们首次证明了 USCOM 可用于测量单心室生理患者的 CI,表明 USCOM 和心导管测量 CI 的一致性可接受,操作者内可重复性高。