Abstract

While scleral lens practise has improved over the years due to factors such as availability of lenses with better materials and designs as well as experience of practitioners, a lack of objectivity appears to remain in terms of assessment of scleral lens fitting. This prospective observational work aimed to achieve standardization on this front through proposing a grading system for scleral lens fitting. After application of prosthetic replacement of ocular surface ecosystem (PROSE) devices on the participants' eyes, four fundamental components for understanding scleral lens fitting such as central and limbal corneal clearance, mid-haptic compression, and alignment of lens edge over anterior sclera were assessed through a series of slit-lamp biomicroscopy imaging as well as with anterior segment optical coherence tomography. FitConnect® was used to modify the device parameters to simulate different grading patterns on the proposed scale. Serial imaging was done for all the different lenses to compose the grading scale. A clinically relevant grading scale was constructed that pictorially demonstrated grades for the different aspect of scleral lens fitting. The grades were conveniently scaled within three categories: "optimal", "acceptable" and "not acceptable". The gradation of scleral lens fitting parameters would take a step towards objectifying the assessment patterns in practise. This will also help reducing the gap between a novice and an experienced practitioner in terms of understanding of scleral lens fitting.

摘要

尽管近年来由于镜片材料和设计的改进以及从业者经验的增加等因素,巩膜镜的实践得到了改善,但在巩膜镜适配评估方面似乎仍然缺乏客观性。本前瞻性观察性研究旨在通过提出巩膜镜适配分级系统来实现这方面的标准化。在参与者的眼睛上应用眼表面修复生态系统(PROSE)设备后,通过一系列裂隙灯生物显微镜成像以及眼前节光学相干断层扫描,评估了理解巩膜镜适配的四个基本要素,如中央和角巩膜间隙、中压区压缩和镜片边缘在巩膜前的对齐情况。使用 FitConnect®修改设备参数,以模拟拟议量表上的不同分级模式。对所有不同的镜片进行连续成像,以组成分级量表。构建了一个具有临床相关性的分级量表,以图形方式展示了巩膜镜适配不同方面的等级。这些等级方便地划分为三个类别:“最佳”、“可接受”和“不可接受”。巩膜镜适配参数的分级将朝着客观评估实践模式迈出一步。这也有助于缩小新手和经验丰富的从业者在理解巩膜镜适配方面的差距。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验