Bergström M, Greitz T, Ribbe T
Neuroradiology. 1986;28(2):100-4. doi: 10.1007/BF00327879.
A method for the determination of stereotaxic coordinates in radiography, e.g. angiography, pneumoencephalography or digital vascular radiography, is described. A special localization frame containing radiopaque structures and scales defines a diagnostic coordinate system. This frame is fixed to the X-ray-table prior to the radiographic procedure and two projections are obtained at arbitrary angles to each other. The focus-film distances do not how to be fixed. The target coordinates are then determined either by a simple graphical procedure or with the use of a digitizing x-y-table, by a computer. With the computer method the films are placed on the digitizing table and the target and a few reference points are marked using a cursor. From the relative positions the computer calculates the coordinates. With the special head fixation system, coordinates of structures visualized in radiographic examinations can be transferred to various therapeutic or diagnostic stereotaxic devices.
描述了一种在放射成像(如血管造影、气脑造影或数字血管造影)中确定立体定向坐标的方法。一种包含不透射线结构和刻度的特殊定位框架定义了一个诊断坐标系。该框架在放射成像程序之前固定在X射线检查台上,并以彼此任意角度获得两个投影。焦点-胶片距离无需固定。然后通过简单的图形程序或使用数字化x-y表由计算机确定目标坐标。使用计算机方法时,将胶片放在数字化表上,并用光标标记目标和几个参考点。计算机根据相对位置计算坐标。借助特殊的头部固定系统,放射成像检查中可视化结构的坐标可转移到各种治疗或诊断立体定向设备上。