Beaumont Health System, Royal Oak, Michigan, USA.
Beaumont Health System, Troy, Michigan, USA.
J Appl Clin Med Phys. 2022 Apr;23(4):e13574. doi: 10.1002/acm2.13574. Epub 2022 Mar 2.
Primary barrier design for linac shielding depends very sensitively on tenth value layer (TVL) data. Inaccuracies can lead to large discrepancies between measured and calculated values of the barrier transmission. Values of the TVL for concrete quoted in several widely used standard references are substantially different than those calculated more recently. The older standard TVL data predict significantly lower radiation levels outside primary barriers than the more recently calculated values under some circumstances. The difference increases with increasing barrier thickness and energy, and it can be as large as a factor of 4 for 18 MV and concrete thickness of 200 cm. This may be due to significant differences in the beam spectra between the earlier and the more recent calculations. Measured instantaneous air kerma rates sometimes show large variations for the same energy and thickness. This may be due to confounding factors such as extra material on, or inside the barrier, variable field size at the barrier, density of concrete, and distal distance from the barrier surface. In some cases, the older TVL data significantly underestimate measured instantaneous air kerma rates, by up to a factor of 3, even when confounding factors are taken into account. This could lead to the necessity for expensive remediation. The more recent TVL values tend to overestimate the measured instantaneous dose rates. Reference TVL data should be computed in a manner that is mathematically consistent with their use in the calculation of air kerma rate outside barriers directly from the linac "dose" rate in MU/min.
直线加速器屏蔽的初级屏蔽设计非常敏感地依赖于十分之一值层 (TVL) 数据。不准确会导致屏蔽传输的测量值和计算值之间出现很大差异。在几个广泛使用的标准参考中引用的混凝土 TVL 值与最近计算的值有很大不同。在某些情况下,旧标准 TVL 数据预测的初级屏蔽外的辐射水平明显低于最近计算的值。随着屏蔽厚度和能量的增加,差异会增大,对于 18 MV 和 200 cm 厚的混凝土,差异最大可达 4 倍。这可能是由于早期和最近计算之间的束谱存在显著差异。对于相同的能量和厚度,瞬时空气比释动能率的测量值有时会有很大差异。这可能是由于屏蔽上或内部的额外材料、屏蔽处的可变射野大小、混凝土密度以及离屏蔽表面的距离等混杂因素造成的。在某些情况下,即使考虑了混杂因素,旧的 TVL 数据也会大大低估测量的瞬时空气比释动能率,最高可达 3 倍。这可能导致需要昂贵的补救措施。最近的 TVL 值往往会高估测量的瞬时剂量率。参考 TVL 数据的计算应与它们在直接从直线加速器“剂量”率(以 MU/min 为单位)计算屏蔽外空气比释动能率的计算中的使用方式在数学上保持一致。