Helsingius P, Hemmilä I, Lövgren T
Clin Chem. 1986 Sep;32(9):1767-9.
We describe a time-resolved fluoroimmunoassay for the cardiac glycoside digoxin. The assay depends on the competitive distribution of Eu3+-labeled anti-digoxin antibodies between solid-phase-bound digoxin and the digoxin in the sample or standard. After this immunoreaction, the bound fraction of the Eu3+-label is dissociated from the solid phase, converted into a highly fluorescent beta-diketone chelate, and measured in a fluorometer with time-resolution (1230 Arcus, LKB-Wallac). Sample volume is 20 microL, the incubation lasts 30 min, and the concentration range of the assay for digoxin is from 0.25 to 4.0 ng/mL (0.32-5.1 nmol of digoxin per liter). Correlation with an in-house RIA was good (r = 0.97, n = 43). The sensitivity was equivalent to that of most RIAs: 0.2 ng/mL (0.25 nmol/L).
我们描述了一种用于检测强心苷地高辛的时间分辨荧光免疫测定法。该测定法基于铕(Eu3+)标记的抗地高辛抗体在固相结合的地高辛与样品或标准品中的地高辛之间的竞争性分布。免疫反应后,铕(Eu3+)标记物的结合部分从固相解离,转化为高荧光的β-二酮螯合物,并在具有时间分辨功能的荧光计(1230 Arcus,LKB-沃拉)中进行测量。样品体积为20微升,孵育持续30分钟,该测定法检测地高辛的浓度范围为0.25至4.0纳克/毫升(每升0.32至5.1纳摩尔地高辛)。与内部放射免疫测定法(RIA)的相关性良好(r = 0.97,n = 43)。灵敏度与大多数放射免疫测定法相当:0.2纳克/毫升(0.25纳摩尔/升)。