Suppr超能文献

一种用于收集嗅裂黏液的新型离心提取装置的研发。

Development of a novel centrifugal extraction device to collect the olfactory cleft mucus.

作者信息

Wu Dawei, Hong Junsheng, Chang Feifan, Wei Yongxiang

机构信息

Department of Otolaryngology, Beijing Anzhen Hospital, Capital Medical University, Beijing, PR China.

Otolaryngology, Beijing Institute of Heart Lung and Blood Vessel Diseases, Beijing, PR China.

出版信息

Acta Otolaryngol. 2022 Mar-Apr;142(3-4):323-328. doi: 10.1080/00016489.2022.2047223. Epub 2022 Mar 15.

Abstract

BACKGROUND

Measuring olfactory cleft mucus biomarkers provides critical insights into the pathogenesis of the olfactory dysfunction.

OBJECTIVES

This study aimed to develop a novel centrifugal extraction device to collect the olfactory cleft mucus.

MATERIAL AND METHODS

The olfactory cleft mucus was collected from 18 patients with chronic rhinosinusitis (CRS) and 10 healthy controls using polyvinyl alcohol (PVA) sponges. Both novel centrifugal extraction devices and traditional devices which consisted of a falcon tube and a syringe shaft were utilized. Levels of Galectin-10 were assessed using the enzyme-linked immunosorbent assay. The olfaction was evaluated with Sniffin' Sticks.

RESULTS

The average extraction efficiency of recovered volume for novel centrifugal extraction devices and traditional devices was 85.44 and 79.15%, respectively. Galectin-10 levels of the olfactory cleft mucus in patients with CRS were significantly higher than that in controls when extracted through novel devices ( < .001). Furthermore, levels of Galectin-10 in patients with CRS significantly correlated with TDI score ( = .020,  = -0.546). There was no significant difference between novel and traditional devices regarding Galectin-10 levels in CRS ( = .966).

CONCLUSIONS AND SIGNIFICANCE

The novel centrifugal extraction device with PVA sponges proved to be a feasible method to collect the olfactory cleft mucus.

摘要

背景

测量嗅裂黏液生物标志物可为嗅觉功能障碍的发病机制提供关键见解。

目的

本研究旨在开发一种新型离心提取装置来收集嗅裂黏液。

材料与方法

使用聚乙烯醇(PVA)海绵从18例慢性鼻-鼻窦炎(CRS)患者和10名健康对照者中收集嗅裂黏液。同时使用了新型离心提取装置和由离心管和注射器轴组成的传统装置。采用酶联免疫吸附测定法评估半乳糖凝集素-10的水平。使用嗅觉棒评估嗅觉。

结果

新型离心提取装置和传统装置回收体积的平均提取效率分别为85.44%和79.15%。通过新型装置提取时,CRS患者嗅裂黏液中的半乳糖凝集素-10水平显著高于对照组(<0.001)。此外,CRS患者的半乳糖凝集素-10水平与TDI评分显著相关(=0.020,=-0.546)。在CRS患者中,新型装置和传统装置在半乳糖凝集素-10水平方面无显著差异(=0.966)。

结论与意义

使用PVA海绵的新型离心提取装置被证明是收集嗅裂黏液的一种可行方法。