Suppr超能文献

利多卡因游离浓度的测定。

Measurement of lidocaine free concentration.

作者信息

Beach C L, Ludden T M, Clementi W A, Allerheiligen S R

出版信息

Ther Drug Monit. 1986;8(3):326-30. doi: 10.1097/00007691-198609000-00015.

Abstract

Since lidocaine exhibits significant variation in serum protein binding, the availability of a practical method for measuring free lidocaine concentration could contribute to the optimization of individual lidocaine dosage regimens. Fifty serum samples from patients receiving lidocaine were partitioned by ultrafiltration and equilibrium dialysis. The lidocaine concentration in the ultrafiltrate was measured using an enzyme multiplied immunoassay (EMIT) and a gas-liquid chromatographic assay (GLC). The lidocaine concentrations in dialysates and filtered retentates were measured by EMIT. Ultrafiltrate concentrations measured by EMIT correlated well with those measured by GLC (r2 = 0.77), but the EMIT results were approximately 10-20% higher than the GLC measurements (GLC = 0.09 + 0.79 EMIT). At least a portion of this difference could be attributed to minor calibrator differences. The concentrations in dialysate and filtered retentate agreed well (r2 = 0.93; filtered retentate = -0.05 + 1.12 X dialysate). The fraction free values obtained by ultrafiltration were slightly lower than those obtained by equilibrium dialysis (0.301 +/- 0.086 vs. 0.345 +/- 0.137; p less than 0.05). It can be concluded that sample partitioning with ultrafiltration and measurement of free lidocaine concentration by EMIT yields results similar to those obtained by equilibrium dialysis or a GLC assay procedure.

摘要

由于利多卡因在血清蛋白结合方面表现出显著差异,因此一种测量游离利多卡因浓度的实用方法的出现可能有助于优化个体利多卡因给药方案。对50份接受利多卡因治疗患者的血清样本进行超滤和平衡透析分离。使用酶放大免疫分析(EMIT)和气液色谱分析(GLC)测量超滤液中的利多卡因浓度。通过EMIT测量透析液和过滤后的保留液中的利多卡因浓度。通过EMIT测量的超滤液浓度与通过GLC测量的浓度相关性良好(r2 = 0.77),但EMIT结果比GLC测量值高约10 - 20%(GLC = 0.09 + 0.79 EMIT)。这种差异至少有一部分可归因于校准物的微小差异。透析液和过滤后的保留液中的浓度一致性良好(r2 = 0.93;过滤后的保留液 = -0.05 + 1.12×透析液)。通过超滤获得的游离分数值略低于通过平衡透析获得的值(0.301±0.086对0.345±0.137;p < 0.05)。可以得出结论,用超滤进行样本分离并通过EMIT测量游离利多卡因浓度所得到的结果与通过平衡透析或GLC分析方法得到的结果相似。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验