To compare the results of computed tomography angiography (CTA), transesophageal echocardiography (TEE), and digital subtraction angiography (DSA) measurements and analyze their accuracy, correlation, and consistency in patients who have successfully undergone left atrial appendage closure (LAAC).
A total of 157 non-valvular atrial fibrillation (AF) patients who underwent LAAC with Watchman devices were included in the study. The maximum diameter and depth of LAA were recorded using CTA, TEE, and DSA. Correlations and agreements were compared.
The LAAC procedure was performed successfully in all patients using the Watchman device. There was no significant difference between DSA and TEE measurements of the diameter of the LAA ostium. LAA ostium diameter obtained by CTA, however, was greater than that from DSA and TEE. Correlations were good between LAA ostium diameter measured by TEE, CTA, and DSA and Watchman device size. DSA measurements and actual device size showed the widest limits of agreement, followed by TEE; CTA measurements showed the narrowest limits of agreement. For LAA depth measurements, mean CTA measurements were higher than those of TEE and DSA. There was no significant difference in depth measurements among the three imaging modalities.
CTA, TEE, and DSA measurements exhibited good correlations with Watchman device size. The ostium diameter and depth of the LAA measured by CTA were greater than those measured by TEE and DSA. The relevance and concordance of CTA measurements were the strongest.
比较计算机断层血管造影(CTA)、经食管超声心动图(TEE)和数字减影血管造影(DSA)测量结果,并分析在成功接受左心耳封堵(LAAC)的患者中它们的准确性、相关性和一致性。
共纳入 157 例接受 Watchman 装置行 LAAC 的非瓣膜性心房颤动(AF)患者。使用 CTA、TEE 和 DSA 记录 LAA 的最大直径和深度。比较相关性和一致性。
所有患者均成功使用 Watchman 装置完成 LAAC 手术。DSA 和 TEE 测量的 LAA 口直径无显著差异。然而,CTA 测量的 LAA 口直径大于 DSA 和 TEE。TEE、CTA 和 DSA 测量的 LAA 口直径与 Watchman 装置大小之间的相关性良好。DSA 测量值与实际装置尺寸显示出最宽的一致性范围,其次是 TEE;CTA 测量值显示出最窄的一致性范围。对于 LAA 深度测量,平均 CTA 测量值高于 TEE 和 DSA。三种成像方式之间深度测量无显著差异。
CTA、TEE 和 DSA 测量值与 Watchman 装置大小具有良好的相关性。CTA 测量的 LAA 口直径和深度大于 TEE 和 DSA 测量值。CTA 测量的相关性和一致性最强。