Department of Cardiovascular and Thoracic Surgery, Faculty of Medicine and Graduate School of Medicine, Hokkaido University, Kita-15, Nishi-7, Kita-ku, Sapporo, 060-8638, Japan.
Gen Thorac Cardiovasc Surg. 2022 Jul;70(7):677-679. doi: 10.1007/s11748-022-01812-8. Epub 2022 Apr 7.
Because a mechanical aortic valve is a contraindication for the implantation of left ventricular assist device, complicated additional procedures such as a replacement with a bioprosthesis and a closure of the left ventricular outflow tract are required to implant the device. Among such procedures, a sandwich plug technique using vascular clips is one of the simple and feasible procedures. However, this technique requires an off-label use of vascular clips within the aorta that could be associated with a risk of dislodgement and embolization. Thus, we developed a modified sandwich technique without using vascular clips, where the valve leaflets were fixed in the closed position using felt patches and sutures instead of vascular clips. This modified technique is a simple and secure method to close the mechanical aortic valve with the minimum use of artificial materials.
由于机械主动脉瓣是植入左心室辅助装置的禁忌症,因此需要进行复杂的附加手术,如更换生物瓣和关闭左心室流出道。在这些手术中,使用血管夹的三明治塞技术是一种简单可行的方法。然而,这种技术需要在主动脉内进行标签外使用血管夹,这可能会增加移位和栓塞的风险。因此,我们开发了一种改良的三明治技术,不使用血管夹,而是使用毛毡补丁和缝线将瓣膜固定在关闭位置,而不是血管夹。这种改良技术是一种简单而安全的方法,可以使用最少的人工材料来关闭机械主动脉瓣。