安布持续气道正压通气(CPAP)系统和Servo 900C呼吸机在CPAP模式下的压力特性

Pressure characteristics of the Ambu CPAP system and the Servo ventilator 900C in CPAP mode.

作者信息

Zetterström H, Jonsson L O

出版信息

Acta Anaesthesiol Scand. 1987 Jan;31(1):104-10. doi: 10.1111/j.1399-6576.1987.tb02530.x.

Abstract

Spontaneous breathing was stimulated in the Ambu continuous positive airway pressure (CPAP) system and the Servo Ventilator 900C by means of a lung model programmed to mimic the respiratory flow patterns of a healthy volunteer and a patient in severe respiratory distress. Changes in airway pressure, flow and volume were recorded during "breathing" with CPAP at 0.5, 1.0 and 1.5 kPa. In the Ambu system, the airway pressure decreased during inspiration and increased during expiration, while the mean airway pressure was close to the pre-set CPAP value. The pressure changes were minimal when the fresh gas flow was increased from 15 to 25 1 X min-1. The higher fresh gas flow is recommendable during deep or rapid breathing. In the Servo ventilator 900C, there was a short initial inspiratory pressure drop, succeeded by a pressure rise above the CPAP value. The expiratory airway pressure was somewhat higher than CPAP. Both systems were found to be recommendable for clinical use.

摘要

通过一个编程以模拟健康志愿者和严重呼吸窘迫患者呼吸气流模式的肺模型,在安必连续气道正压通气(CPAP)系统和Servo Ventilator 900C中刺激自主呼吸。在使用0.5、1.0和1.5 kPa的CPAP进行“呼吸”期间,记录气道压力、流量和容积的变化。在安必系统中,气道压力在吸气时降低,在呼气时升高,而平均气道压力接近预设的CPAP值。当新鲜气体流量从15升/分钟增加到25升/分钟时,压力变化最小。在深呼吸或快速呼吸期间,建议采用较高的新鲜气体流量。在Servo Ventilator 900C中,最初有一个短暂的吸气压力下降,随后压力上升超过CPAP值。呼气气道压力略高于CPAP。发现这两种系统都推荐用于临床。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验