Martin R G, Safir A
J Cataract Refract Surg. 1987 Jan;13(1):62-5. doi: 10.1016/s0886-3350(87)80014-7.
Asteroid hyalosis can create ultrasonic echoes capable of interfering with A-scan ultrasonic measurement of ocular length. Two cases are reported, one of which includes a measurement error necessitating a second implantation. Surgeons who entrust measurements to technicians should be aware of the measurement technique and should establish protocols that preclude such errors.