Piedmont Heart Institute, Suite 2065, 95 Collier Road, Atlanta, GA 30309, USA.
Biotronik, Inc., 6024 Jean Road, Lake Oswego, OR 97035, USA.
Cardiovasc Revasc Med. 2022 Oct;43:97-101. doi: 10.1016/j.carrev.2022.04.009. Epub 2022 Apr 25.
BACKGROUND/PURPOSE: The PK Papyrus covered coronary stent system (Biotronik AG, Bülach, Switzerland) is intended for treatment of coronary artery perforation (CAP) and is approved for use under a Humanitarian Device Exemption (HDE) in the United States (US).
METHODS/MATERIALS: The retrospective data analysis includes cases reported from the US PK Papyrus HDE post-market surveillance clinical dataset with CAP cited as the reason for PK Papyrus stent use.
From April 2019 to July 2021, PK Papyrus device registration forms citing CAP as the reason for use were received for 1094 cases from 335 US hospital programs. Ellis classification was assessed as: type III cavity spilling/IV, 11.0%; type III, 57.9%; type II, 23.8%; type I, 7.3%. Mechanisms of perforation included: balloon angioplasty (42.3%), stent placement (31.3%), atherectomy (13.9%), and guidewire (10.9%). The majority (72.6%) of cases involved single covered stent placement. Successful PK Papyrus delivery was reported in 97.7% of cases with successful perforation sealing in 92.1%. Emergency cardiac surgery and in-hospital death occurred in 6.3% and 12.4% of cases, respectively. Pericardiocentesis was performed in 30.2% of patients. Acute/subacute stent thrombosis occurred in 10 patients (1.1%).
As the largest dataset of patients treated with a covered stent for CAP, these data provide significant insight into patient characteristics, procedural outcome, and in-hospital clinical events associated with this life-threatening complication. These results demonstrate that the PK Papyrus stent is a safe and effective method to seal CAP and with the potential to reduce high morbidity and mortality associated with this event.
背景/目的:PK Papyrus 覆盖式冠状动脉支架系统(Biotronik AG,瑞士比拉赫)旨在治疗冠状动脉穿孔(CAP),并已获得美国(美国)人类设备豁免(HDE)的批准使用。
方法/材料:回顾性数据分析包括从美国 PK Papyrus HDE 上市后监测临床数据集报告的病例,CAP 被列为使用 PK Papyrus 支架的原因。
从 2019 年 4 月至 2021 年 7 月,从 335 个美国医院项目收到了 1094 例 PK Papyrus 设备登记表,其中 CAP 被列为使用原因。Ellis 分类评估为:III 型腔溢出/IV 型,11.0%;III 型,57.9%;II 型,23.8%;I 型,7.3%。穿孔机制包括:球囊血管成形术(42.3%)、支架放置(31.3%)、旋切术(13.9%)和导丝(10.9%)。大多数(72.6%)病例涉及单个覆盖式支架放置。97.7%的病例成功交付 PK Papyrus,92.1%的病例成功密封穿孔。紧急心脏手术和院内死亡分别发生在 6.3%和 12.4%的病例中。30.2%的患者进行了心包穿刺术。10 名患者(1.1%)发生急性/亚急性支架血栓形成。
作为治疗 CAP 的覆盖式支架患者的最大数据集,这些数据提供了对患者特征、手术结果和与这种危及生命的并发症相关的院内临床事件的重要见解。这些结果表明,PK Papyrus 支架是一种安全有效的 CAP 密封方法,有可能降低与该事件相关的高发病率和死亡率。