J Xray Sci Technol. 2022;30(4):689-696. doi: 10.3233/XST-221181.
To compare image quality and radiation dose of computed tomography angiography (CTA) of the head and neck in patients using two Gemstone Spectral Imaging (GSI) scanning protocols.
A total of 100 patients who underwent head-neck CTA were divided into two groups (A and B) according to the scanning protocols, with 50 patients in each group. The patients in group A underwent GSI scanning protocol 1 (GSI profile: head and neck CTA), while those in group B underwent GSI scanning protocol 2 (GSI profile: chest 80 mm). All images were reconstructed using 40% and 70% pre- and post-adaptive level statistical iterative reconstruction V (pre-ASiR-V and post-ASiR-V) algorithms, respectively. The CT dose index (CTDIvol) and dose-length (DLP) product were recorded and the mean value was calculated and converted to the effective dose. CT values, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) of all images were calculated. Additionally, subjective image evaluation was conducted by two independent radiologists using a five-point scoring method. All data were statistically analyzed.
There were no significant differences in the CT values, SNR, CNR, and subjective score between groups A and B (p > 0.05); however, the mean effective dose (1.2±0.1 mSv) in group B was 45.5% lower than that in group A (2.2±0.2 mSv) (p < 0.05).
GSI scanning protocol 2 could more effectively reduce the radiation dose in head-neck CT angiography while maintaining image quality compared to GSI scanning protocol 1.
比较两种宝石能谱成像(GSI)扫描方案在头颈部 CT 血管造影(CTA)中患者的图像质量和辐射剂量。
将 100 例行头颈部 CTA 的患者按扫描方案分为两组(A 组和 B 组),每组 50 例。A 组患者采用 GSI 扫描方案 1(GSI 谱:头颈部 CTA),B 组患者采用 GSI 扫描方案 2(GSI 谱:胸部 80mm)。所有图像均采用 40%和 70%预和自适应统计迭代重建 V 级(pre-ASiR-V 和 post-ASiR-V)算法进行重建。记录 CT 剂量指数(CTDIvol)和剂量长度乘积(DLP),并计算平均值并转换为有效剂量。计算所有图像的 CT 值、信噪比(SNR)和对比噪声比(CNR)。此外,两名独立放射科医生采用五分制对图像质量进行主观评价。对所有数据进行统计学分析。
A 组和 B 组间的 CT 值、SNR、CNR 和主观评分均无统计学差异(p>0.05);但 B 组的平均有效剂量(1.2±0.1mSv)比 A 组(2.2±0.2mSv)低 45.5%(p<0.05)。
与 GSI 扫描方案 1 相比,GSI 扫描方案 2 能更有效地降低头颈部 CTA 的辐射剂量,同时保持图像质量。