Moussous Ouiza
Departement of Medical Physics, Nuclear Research Center of Algiers, Algiers, Algeria.
J Med Phys. 2022 Jan-Mar;47(1):105-108. doi: 10.4103/jmp.jmp_83_21. Epub 2022 Mar 31.
The effective atomic number and electron density of some Fricke gel dosimeters were calculated for photon interaction in the energy range from 10 keV to 1000 MeV using Auto-Z, direct and power law methods. The results are presented relative to those of water to allow direct comparison. It is found, that the effective atomic numbers and effective electron densities calculated with the Auto-Z and direct methods, demonstrates a good agreement in the energy interval extending from 0.1 MeV to 10 MeV. For effective atomic number relative to water, Ferrous Agarose Xylenol gel showed better water equivalence with difference up to 0.3%, while FX-PVA-GTA and Ferrous Xylenol Gelatin gels showed differences up to 2.26% and 2.25%, respectively.