定量分析踝足矫形器制作和适配过程中的对线偏差:一项单中心研究。

Quantifying alignment bias during the fabrication and fitting of ankle-foot orthoses: A single center study.

机构信息

Gillette Children's Specialty Healthcare, St. Paul, MN, United States.

Gillette Children's Specialty Healthcare, St. Paul, MN, United States.

出版信息

Gait Posture. 2022 Jul;96:29-34. doi: 10.1016/j.gaitpost.2022.05.007. Epub 2022 May 7.

Abstract

BACKGROUND

The sagittal plane alignment of ankle-foot orthoses (AFO) and AFO footwear combinations (AFO-FC) has been shown to influence gait outcomes. As such, clinicians often target a particular alignment during the fabricating and fitting of an AFO to maximize outcomes.

RESEARCH QUESTION

How does the alignment of an AFO change during the fabrication and fitting process with respect to the intended, benchmark sagittal plane alignment identified by the consulting orthotist?

STUDY DESIGN

Prospective METHODS: The assessment of AFO alignment was performed using a convenience sample of 125 custom molded AFOs from 68 individuals fabricated at our center (57 bilateral AFOs, 11 unilateral AFOs). The alignment of each AFO was measured at 5 distinct steps during the fabrication and fitting process using a recently validated method to measure AFO neutral angle using differential inclinometers.

RESULTS

Prior to fabrication, the intended, benchmark alignment set by the consulting orthotist was 90 degrees for 92% of AFOs, was between 1 and 7 degrees of dorsiflexion for 7% of AFOs and was 5 degrees of plantarflexion for 1% of AFOs. Repeated measures ANOVA showed that AFO alignment changed between all fabrication and fitting steps. Overall, paired t-tests confirmed that AFO alignment was consistently 2-5 degrees more dorsiflexed than the benchmark alignment. Prior to fitting shoes, 55% of fabricated AFOs measured more than 2 degrees from the benchmark alignment. After fitting shoes, nearly 87% of AFO-FCs were more than 2 degrees from the benchmark alignment.

SIGNIFICANCE

The finding of systematic dorsiflexion bias and changes in AFO alignment throughout the fabrication and fitting process indicates the need to improve AFO fabrication precision. The neutral angle measurement methodology - using differential inclinometers - provides a means to improve this precision by enabling orthotists to precisely quantify and make appropriate adjustments to AFO alignment throughout the entire fabrication and fitting process.

摘要

背景

踝足矫形器(AFO)和 AFO 鞋类组合(AFO-FC)的矢状面排列已被证明会影响步态结果。因此,临床医生在制作和适配 AFO 时经常针对特定的排列,以最大限度地提高效果。

研究问题

在制作和适配过程中,AFO 的排列如何相对于由顾问矫形师确定的预期基准矢状面排列发生变化?

研究设计

前瞻性

方法

使用我们中心制作的 125 个定制成型 AFO 的便利样本(68 名个体的 57 个双侧 AFO 和 11 个单侧 AFO)来评估 AFO 对齐情况。使用最近验证的方法,使用差动测斜仪测量 AFO 中立角度,在制作和适配过程中的 5 个不同步骤测量每个 AFO 的对齐情况。

结果

在制作之前,顾问矫形师设定的预期基准排列为 90 度,占 AFO 的 92%,AFO 的背屈角度为 1 至 7 度,占 AFO 的 7%,跖屈角度为 5 度,占 AFO 的 1%。重复测量方差分析显示,AFO 对齐在所有制作和适配步骤之间都发生了变化。总体而言,配对 t 检验证实,AFO 对齐始终比基准对齐背屈 2 至 5 度。在安装鞋子之前,55%的制作好的 AFO 与基准排列相差 2 度以上。安装鞋子后,近 87%的 AFO-FC 与基准排列相差 2 度以上。

意义

发现 AFO 排列在整个制作和适配过程中存在系统性背屈偏差和变化,这表明需要提高 AFO 制作精度。使用差动测斜仪的中立角度测量方法为通过使矫形师能够在整个制作和适配过程中精确量化和进行适当的 AFO 对齐调整,提供了一种提高这种精度的方法。