Characterization of normal and malignant breast tissues using X-ray scattering techniques has shown promising results and applications.
To examine possibility of characterizing normal and malignant breast tissues using the scattered photon distribution of polyenergetic beams of 30 kV X-rays.
A Monte Carlo simulation is upgraded so that it is capable of simulating input mammographic X-ray spectra from different target-filter combinations, tracing photon transport, and producing the distribution of scattered photons. The target-filter combinations include Mo-Mo, Mo-Al, Mo-Rh, Rh-Rh, Rh-Al, W-Rh, and W-Al. Analysis of obtained scattered photon distribution is carried out by comparing the ratio of count under the peak in the momentum transfer region from 0 to 1.55 nm-1, to that in the region from 1.6 to 9.1 nm-1 (covering the regions of scattering from fat and soft tissue, respectively) for breast samples with different percentages of normal tissue (0-100%).
Mo-Mo target-filter combination shows a high linear dependence of the count under peak ratio on the percentage of normal tissue in breast samples (R2 = 0.9513). Despite slightly less linear than Mo-Mo, target-filter combinations other than Rh-Rh, W-Rh, and W-Al produce high linear responses (R2 > 0.9)CONCLUSION:Mo-Mo target-filter combination would probably be the most relevant in characterizing normal and malignant breast tissues from their scattered photon distribution.
使用 X 射线散射技术对正常和恶性乳腺组织进行特征描述已显示出有前景的结果和应用。
研究使用 30kV 多能量 X 射线的散射光子分布来描述正常和恶性乳腺组织的可能性。
升级蒙特卡罗模拟,使其能够模拟来自不同靶-滤组合的乳腺 X 射线光谱,追踪光子传输,并产生散射光子的分布。靶-滤组合包括 Mo-Mo、Mo-Al、Mo-Rh、Rh-Rh、Rh-Al、W-Rh 和 W-Al。通过比较动量传递区域(0 至 1.55nm-1)下的计数与脂肪和软组织散射区域(分别为 1.6 至 9.1nm-1)下的计数的比值,分析获得的散射光子分布。对不同正常组织百分比(0-100%)的乳腺样本进行分析。
Mo-Mo 靶-滤组合显示出乳腺样本中正常组织百分比与峰值下计数比值之间具有很高的线性相关性(R2=0.9513)。尽管不如 Mo-Mo 线性,但 Mo-Rh、W-Rh 和 W-Al 以外的靶-滤组合产生了较高的线性响应(R2>0.9)。
Mo-Mo 靶-滤组合可能是从散射光子分布特征描述正常和恶性乳腺组织的最相关组合。