Haacke E M, Lenz G W, Nelson A D
Magn Reson Med. 1987 Feb;4(2):162-74. doi: 10.1002/mrm.1910040209.
This note explains and illustrates a technique of reducing artifacts produced by periodic motion without using physiologic gating. The method is simple and can be applied on any standard MRI unit. For periodic motion, effective gating can be attained by setting the product of the number of acquisitions (at each phase-encoding step), N, times the repeat time, TR, equal to the period of the motion, T. This pseudo-gating can be used with any TR but, if changes in the period occur, is most robust with short TR values. The method is also applicable to multislice and volume imaging. For fast field echo methods, the above rule can be used if the period of the motion is smaller than the total scan time. Otherwise, the scan need only be repeated N times to avoid artifacts and improve signal-to-noise. The method has been implemented to remove both respiratory and/or cardiac motion artifacts.
本注释解释并说明了一种在不使用生理门控的情况下减少周期性运动产生的伪影的技术。该方法简单,可应用于任何标准MRI设备。对于周期性运动,通过将(在每个相位编码步骤)采集次数N与重复时间TR的乘积设置为等于运动周期T,可实现有效的门控。这种伪门控可与任何TR一起使用,但如果运动周期发生变化,在短TR值时最为稳健。该方法也适用于多层和容积成像。对于快速场回波方法,如果运动周期小于总扫描时间,则可使用上述规则。否则,只需重复扫描N次以避免伪影并提高信噪比。该方法已用于去除呼吸和/或心脏运动伪影。