Pankratov M M, Pomerantzeff O, Pflibsen K P, Bearse G R
Ophthalmic Surg. 1987 Jan;18(1):61-5.
A step-zoom probe for laser endophotocoagulation of the eye has been developed, using fiber optics and a gradient index (GRIN) rod lens. It provides the operator with five or more focused spots ranging in size from 50 to 500 microns, and the spot size on the target is independent of the media filling the vitreous cavity, i.e., fluid or air. Each spot is located at the waist of the beam and therefore can be easily placed on the target, reducing the potential for overtreatment. A prototype model is currently undergoing animal testing.