Suppr超能文献

加热丝加湿器加湿性能与电加热片温度监测

Monitoring of Heated Wire Humidifier Hygrometric Performances With Heater Plate Temperature.

机构信息

Centre de Recherche de l'Institut Universitaire de Cardiologie et de Pneumologie de Québec, Québec, Québec, Canada.

出版信息

Respir Care. 2022 Sep;67(9):1147-1153. doi: 10.4187/respcare.10014. Epub 2022 Jun 7.

Abstract

BACKGROUND

Under-humidification and associated complications may occur with heated humidifiers. Hygrometric performances of heated wire humidifiers are reduced by high ambient and high outlet ventilator temperatures. Currently, there is no reliable monitoring tool to evaluate humidification performances of heated wire humidifiers in the daily practice. We sought to demonstrate the relation between humidity delivered by heated wire humidifiers and different parameters that could be used to monitor humidity of gas delivered to subjects.

METHODS

On a bench test, we measured heater plate temperature, inlet chamber temperature, and delivered humidity with MR850 system. Temperature displayed on the humidifier was also recorded. The measurements were performed at different ambient temperatures and five minute ventilation levels (5, 7.5, 10, 12.5, 15 L/min). Inlet chamber temperatures varied from 20-40°C. In each condition, hygrometric measurements with the psychrometric method were performed at steady state.

RESULTS

We performed 279 measurements at steady state including all conditions. We found a good correlation between heater plate temperature and absolute humidity delivered (R = 0.82). This relationship was hardly affected by ambient temperature, but minute ventilation had more effect. For different minute ventilations, the correlation between heater plate temperature and absolute humidity delivered was very good with coefficient of determination R from 0.87-0.98. Heater plate temperature > 62°C was a good predictor of absolute humidity delivered > 30 mg HO/L (area under the curve = 0.96, sensibility 79%, specificity 94%). No correlation existed between humidity delivered and the outlet chamber temperature (displayed on the humidifier).

CONCLUSIONS

In this bench study, we have shown a good correlation between heater plate temperature and humidity delivered with a heated wire humidifier. This means that a "hidden" hygrometer is built into the heated wire humidifier. Heater plate temperature should be used as a surrogate of humidity to improve the humidification monitoring.

摘要

背景

加热加湿器可能会出现湿化不足和相关并发症。高环境温度和高出口通风机温度会降低加热丝加湿器的湿度性能。目前,在日常实践中,没有可靠的监测工具来评估加热丝加湿器的湿化性能。我们旨在证明加热丝加湿器输送的湿度与可用于监测输送给受试者的气体湿度的不同参数之间的关系。

方法

在台架测试中,我们使用 MR850 系统测量了加热丝加湿器的加热板温度、进气室温度和输送湿度。还记录了加湿器上显示的温度。在不同的环境温度和 5 分钟通气水平(5、7.5、10、12.5、15 L/min)下进行了测量。进气室温度在 20-40°C 之间变化。在每种条件下,在稳定状态下使用干湿球法进行了湿度测量。

结果

我们在稳定状态下共进行了 279 次测量,包括所有条件。我们发现加热板温度与输送的绝对湿度之间存在良好的相关性(R = 0.82)。这种关系几乎不受环境温度的影响,但通气量的影响更大。对于不同的通气量,加热板温度与输送的绝对湿度之间的相关性非常好,决定系数 R 从 0.87-0.98 不等。加热板温度>62°C 是输送的绝对湿度>30 mg HO/L 的良好预测指标(曲线下面积=0.96,灵敏度 79%,特异性 94%)。输送的湿度与出气室温度(显示在加湿器上)之间不存在相关性。

结论

在这项台架研究中,我们已经证明了加热丝加湿器的加热板温度与输送的湿度之间存在良好的相关性。这意味着加热丝加湿器内置了一个“隐藏”的湿度计。应将加热板温度用作湿度的替代指标,以改善湿化监测。