Clinic for Oral Health & Medicine, University Center for Dental Medicine Basel UZB, University of Basel, Basel, Switzerland.
Center of Biomechanics and Biocalorimetry, c/o Department of Biomedical Engineering (DBE), University of Basel, Allschwil, Switzerland.
PLoS One. 2022 Jun 10;17(6):e0269600. doi: 10.1371/journal.pone.0269600. eCollection 2022.
Differential scanning calorimetry (DSC) has been used widely to study various biomarkers from blood, less is known about the protein profiles from saliva. The aim of the study was to investigate the use DSC in order to detect saliva thermal profiles and determine the most appropriate sampling procedure to collect and process saliva. Saliva was collected from 25 healthy young individuals and processed using different protocols based on centrifugation and filtering. The most effective protocol was centrifugation at 5000g for 10 min at 4°C followed by filtration through Millex 0.45 μm filter. Prepared samples were transferred to 3 mL calorimetric ampoules and then loaded into TAM48 calibrated to 30°C until analysis. DSC scans were recorded from 30°C to 90°C at a scan rate of 1°C/h with a pre-conditioning the samples to starting temperature for 1 h. The results show that the peak distribution of protein melting points was clearly bimodal, and the majority of peaks appeared between 40-50°C. Another set of peaks is visible between 65°C- 75°C. Additionally, the peak amplitude and area under the peak are less affected by the concentration of protein in the sample than by the individual differences between people. In conclusion, the study shows that with right preparation of the samples, there is a possibility to have thermograms of salivary proteins that show peaks in similar temperature regions between different healthy volunteers.
差示扫描量热法(DSC)已广泛用于研究来自血液的各种生物标志物,而对唾液中的蛋白质谱了解较少。本研究旨在探讨使用 DSC 来检测唾液热谱,并确定收集和处理唾液的最合适采样程序。从 25 名健康年轻个体中收集唾液,并使用基于离心和过滤的不同方案进行处理。最有效的方案是在 4°C 下以 5000g 离心 10 分钟,然后通过 0.45μm Millex 过滤器过滤。将制备好的样品转移到 3mL 量热安瓿中,然后加载到 TAM48 中,校准至 30°C 进行分析。以 1°C/h 的扫描速率从 30°C 记录到 90°C 的 DSC 扫描,将样品预温至起始温度 1 小时。结果表明,蛋白质熔点的峰分布明显呈双峰型,大多数峰出现在 40-50°C 之间。另一组峰可见于 65°C-75°C 之间。此外,与样品中蛋白质的浓度相比,峰的幅度和峰下面积受个体间差异的影响较小。总之,该研究表明,通过正确制备样品,有可能获得显示不同健康志愿者之间在相似温度区域出现峰的唾液蛋白质热谱。