Laboratory of Haematology and Blood Bank Unit, "Attiko" Hospital, School of Medicine, National and Kapodistrian University of Athens, Athens, Greece.
Neonatal Intensive Care Unit, "Agios Panteleimon" General Hospital of Nikea, Piraeus, Greece.
Eur J Haematol. 2022 Oct;109(4):327-335. doi: 10.1111/ejh.13812. Epub 2022 Jun 22.
To investigate the agreement between the EXTEM and NATEM measurements.
In this retrospective observational study, EXTEM and NATEM analyses were performed on blood samples from 162 ill neonates, providing 324 paired measurements. The agreement between EXTEM and NATEM measurements was evaluated by the nonparametric spearman's rank correlation to assess the correlation between the paired measurements, by the Bland-Altman analysis for the graphical presentation of the agreement, and by the Deming regression model to assess the significance of the agreement. The agreement between the two methods for the detection of bleeding events was determined by kappa statistic.
Strong correlations were found between EXTEM and NATEM measurements for A10, MCF. The Bland-Altman plots showed good agreement for A10, MCF, LI60, and alpha angle parameters, while CT showed a nearly linear slope indicating that bias increased with the mean. The highest agreement for bleeding events was found for the A10 parameter (κ = 0.70, p < .001), while the lowest for the CT parameter (κ = 0.36, p = .94).
NATEM parameters that reflect clot firmness and fibrinolytic activity are strongly correlated with the corresponding EXTEM measurements with a good agreement between them, indicating that these two methods could be used interchangeably.
研究 EXTEM 和 NATEM 测量值之间的一致性。
在这项回顾性观察研究中,对 162 名患病新生儿的血液样本进行了 EXTEM 和 NATEM 分析,共提供了 324 对测量值。通过非参数 Spearman 秩相关评估了 EXTEM 和 NATEM 测量值之间的一致性,以评估配对测量值之间的相关性;通过 Bland-Altman 分析以图形方式展示一致性;通过 Deming 回归模型评估一致性的显著性。通过 Kappa 统计评估了两种方法检测出血事件的一致性。
A10、MCF 等 EXTEM 和 NATEM 测量值之间存在很强的相关性。Bland-Altman 图显示 A10、MCF、LI60 和 alpha 角参数的一致性较好,而 CT 则显示出近乎线性的斜率,表明随着均值的增加,偏差也随之增加。A10 参数对出血事件的一致性最高(κ=0.70,p<0.001),而 CT 参数的一致性最低(κ=0.36,p=0.94)。
反映血凝块坚固性和纤维蛋白溶解活性的 NATEM 参数与 EXTEM 测量值高度相关,且两者之间具有良好的一致性,这表明这两种方法可以互换使用。