Suppr超能文献

采用 LC-ESI-MS/MS 分析 B 族维生素的混合表面技术评价。

Evaluation of hybrid surface technology for the analysis of the B-group vitamins by LC-ESI-MS/MS.

机构信息

Waters Corporation, Milford, MA 01757, USA.

Department of Metabolism, Digestion and Reproduction, Faculty of Medicine, Imperial College, London SW7 2AZ, UK.

出版信息

J Chromatogr B Analyt Technol Biomed Life Sci. 2022 Jul 15;1204:123336. doi: 10.1016/j.jchromb.2022.123336. Epub 2022 Jun 11.

Abstract

Recently, a novel hybrid surface technology (HST) has been developed to mitigate metal analyte adsorption in liquid chromatography. The HST provides a hybrid organic-inorganic surface on the metal fluidic path, from injection to detector and including the column frits and wall, to mitigate the interaction between analytes and metals. Here the impact of the HST on the analysis of B group vitamins using liquid chromatography coupled with electrospray tandem mass spectrometry (LC-ESI-MS/MS) has been evaluated. Significant improvements in analyte intensity, limit of quantification (LOQ), carry-over, and peak shape were observed using an LC-ESI-MS/MS system and column that incorporated the HST. The key observed improvements include a 3-10 times increase in sensitivity (providing a lower LOQ) for riboflavin, thiamine, nicotinamide, FMN, PLP, and 5MTHF, no carry-over, and a more symmetrical peak for thiamine. When applied to the analysis of B group vitamins in energy drinks and B vitamin dietary supplement samples, the HST system demonstrated excellent accuracy and repeatability.

摘要

最近,开发了一种新型混合表面技术(HST),以减轻液相色谱中金属分析物的吸附。HST 在金属流路(从进样到检测器,包括柱接头和管壁)上提供了一种混合有机-无机表面,以减轻分析物与金属之间的相互作用。本文评估了 HST 对使用液相色谱-电喷雾串联质谱(LC-ESI-MS/MS)分析 B 族维生素的影响。使用 LC-ESI-MS/MS 系统和包含 HST 的色谱柱,观察到分析物强度、定量限(LOQ)、交叉污染和峰形有显著改善。观察到的主要改进包括核黄素、硫胺素、烟酰胺、FMN、PLP 和 5MTHF 的灵敏度提高 3-10 倍(提供更低的 LOQ),无交叉污染,以及硫胺素的峰形更对称。当应用于能量饮料和 B 族维生素膳食补充剂样品中 B 族维生素的分析时,HST 系统表现出优异的准确性和重复性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验